Maintenance
“Not Adequate” message
Arc power and splicing position calibration are completed but further calibration is
strongly recommended, as the change from the previous arc calibration is too large.
Press ENT key to perform arc calibration, or ESC key to exit even though arc calibration
is not completed.
Motor Calibration
Motors were adjusted at the factory before shipping. However, settings could change due to
various reasons. This function automatically calibrates the speed of all motors.
Operation Procedure
1. Select the [Motor Calibration] in the [Maintenance Menu].
2. Load prepared fibers in the splicer and press ENT key.
3. Speeds for all motors are automatically calibrated.
In some cases, multiple iterations of arc calibration are needed until the calibration
process is successfully completed and the "Test Finish" message is displayed. Arc
calibration can be considered almost completed if multiple iterations are completed
without receiving the message.
Number threshold can be set so that "Test Finish" message is displayed after specific
number of arc calibrations is performed. See section [Machine Settings] for detail.
Result: Not Adequate
Perform this function when "Fat" or "Thin" error has occurred.