14
Clinical safety
Electromagnetic compatibility (EMC)
See the safety section of the ultrasound system user guide and supplemental guides for information on
electromagnetic compatibility with IEC 60601-1-2, including the Manufacturer’s Declaration.
Cautions:
Excessive bending or twisting of cables can cause a failure or intermittent
operation.
Improper cleaning or disinfecting of any part of the accessories and peripherals
can cause permanent damage. For cleaning and disinfecting instructions, see
on page 11 and the manufacturer’s operating
instructions.
Use only peripherals recommended by FUJIFILM SonoSite.
Accessible metal on the signal inputs and outputs on the back and side of the
ultrasound system is not protectively earthed. Do not perform a high current
ground impedance test involving this part.
WARNING:
To avoid misdiagnosis, do not use non-medical (commercial) grade peripheral
monitors. These monitors have not been verified or validated by FUJIFILM
SonoSite as being suitable for diagnosis.