The value for range 1 and range 2 must fall within
the range from the MIN and the MAX range
(including the MIN and the MAX range), and
at the same time range 1 must be smaller than
range 2.
The number of ranges is 1 or 2.
Settable range
Maintenance
Mode
Range set
Ch1
Ch2
Ch3
Ch4
NOx
SO
2
CO
O
2
Select Ch No.
ENT
Maintenance
Mode
Range set
Ch1 NOx
MIN range
Range 1
Range 2
MAX range
Range num.
500.0
500.0
5000.
5000.
2
ppm
ppm
ppm
ppm
Select range or
range num.
Maintenance
Mode
Range set
Ch1 NOx
MIN range
Range 1
Range 2
MAX range
Range num.
500.0
5
00.0
5000.
5000.
2
ppm
ppm
ppm
ppm
Set range
ENT
Be sure to perform zero / span calibration when the range setting is changed.
Otherwise, the measurement value may not be output properly.
Caution