700-SDI-PT-KEL-Man      Rev 4    06 Jan 2020 

15/19 

 

Offset value for Temperature  

The offset is always given in the selected unit.  The internal transmitter lays it down in 

C and then 

converts to the selected unit.   
 
Default value = 0.000 
Offset value range = +/-2

If an offset outside of the range is programmed, an error occurs and the previous offset value is still 
valid. 
 
The offset is always given in the selected unit.  The internal transducer lays it down in 

C and then 

converts to the selected unit.   
 
The offset value for the temperature is saved in the EEPROM and can be read and changed with the 
command: 
 

Description

 

Command

 

Response

 

Read offset value for temperature TOB 

aXZZTOB!  

a<value><CR><LF> 

Write offset value for temperature TOB

8

  aXZZTOB<value>! 

a<value><CR><LF> 

 

 

Temperature Value 

The internal transducer always reads in °C. It is only multiplied with the “unit”-factor and added with 
its offset before it is given out over SDI-12 communication.  
 
There are three different temperature units available to choose. They can be selected via a table 
which is laid down in the internal EEPROM: 
 

Number   Pressure Unit 

Factor 

Offset 

00 

Factory 

(1) 

01 

02 

1.8001 

32

 

03 

273.15

 

 

Description

 

Command

 

Response

 

read temperature unit 

aXT!  

a0x<CR><LF> 

write temperature unit 

aXT0x! 

a0x<CR><LF> 

 

 

Continuous Mode 

In continuous mode, the time interval and buffer size are defined.  The buffer size defines the 
number of pressure values that are stored.  After the interval time elapses the sensor wakes up 
from low power mode, takes a measurement, and then returns to low power mode. The pressure 
values are buffered in a ring buffer  and it is possible to read out the average value of this buffer 
over an aR0! or aRC0! command.   
 

 

8

 Top of Bridge – method whereby temperature is derived using a Wheatstone bridge 

Содержание SDI-PT-KEL

Страница 1: ...XTREME ENVIRONMENTS EXTREMELY SIMPLE SDI PT KEL SDI 12 Submersible Pressure Transducer User Manual For SDI PT SS KEL and SDI PT TT KEL 1 800 548 4264 www ftsinc com 700 SDI PT KEL Man Rev 4 05 Feb 202...

Страница 2: ...ormation Canadian Headquarters 1065 Henry Eng Place Victoria BC V9B 6B2 Canada www ftsinc com Toll free 1 800 548 4264 Local 250 478 5561 Technical support portal http support ftsinc com Email service...

Страница 3: ...ALLATION WITH AN FTS AXIOM DATALOGGER 5 3 1 GENERAL INFORMATION 6 3 2 NOTATION FOR SDI 12 COMMANDS 6 3 3 ERROR DISPLAYS 7 3 4 MEASUREMENT COMMANDS 8 3 4 1 Concurrent Measurement Commands 9 3 4 2 Conti...

Страница 4: ...with flying leads for use with other data loggers A vent tube incorporated in the cable vents the sensor diaphragm to the atmosphere This eliminates the need to compensate for changes in barometric pr...

Страница 5: ...ysical conditions looking for damage to the sensor and cable or sensor blockages Check the vent tube enclosure humidity monitor card to determine if the desiccant requires changing Verify that the air...

Страница 6: ...ixture of the cleaner and fresh water and the third container with fresh water Hold the cable approximately 6 inches from the transducer and gently stir in each of the containers for 20 30 seconds in...

Страница 7: ...ne with anything Damage to the sensing membrane will require repair or replacement Do not twist crimp or compress the cable as this can damage the pressure tube o The cable should not be bent to a rad...

Страница 8: ...ll or attached to a rigid conduit with a conduit fitting which is integral with the transducer The SDI PT KEL should be installed at a fixed depth mounted vertically below the lowest expected water st...

Страница 9: ...he data This chapter will provide basic direction on general SDI 12 Commands Detailed direction can be found in accordance with the document found at footnote 5 This chapter will also provide informat...

Страница 10: ...eeded The brackets are not included in the command xxx Floating point number Number representing a measurement value or command parameter If a command parameter it will be bolded and may be marked as...

Страница 11: ...he command string will include the CRC cyclic redundancy check Command Response Values Measured send aDO command to retrieve aM aMC a0012 After 1 second 2 measurements available Pressure Temperature a...

Страница 12: ...and C2 commands are always returned in the stated units of measurement regardless of the selected temperature and pressure units 3 4 2 Continuous Measurement R Commands Continuous measurement must be...

Страница 13: ...value returned u uu second measurement value returned u uu etc The values are displayed with a polarity sign followed by a maximum of seven digits which may or may not have a decimal point The maximu...

Страница 14: ...only one SDI device should be connected to the bus when using this command problems arise when several devices respond on the same bus STRING NOTE Cmd Request the single device on this bus to report...

Страница 15: ...SDI 12 commands in that the first character of every command must be a sensor address which is then followed by the X command and terminated by an exclamation mark Likewise the first character of a r...

Страница 16: ...ult 9 80665m s2 aXG aXG value a value CR LF Read write temperature unit aXT aXT0x a0x CR LF Read write offset value for temperature T aXZZT aXZZT value a value CR LF Read write offset value for temper...

Страница 17: ...selected unit Default gain 1 000 Gain range 0 80 1 20 If a gain outside of range is programmed an error occurs and the previous gain value remains valid The gain value for the pressure is saved in the...

Страница 18: ...transducer always reads in C It is only multiplied with the unit factor and added with its offset before it is given out over SDI 12 communication There are three different temperature units available...

Страница 19: ...N aXRS tt B ss RUN CR LF aXRM tt B ss RUN CR LF Stop continuous measurement aXROFF aXRS tt B ss STOP CR LF aXRM tt B ss STOP CR LF Example To set and start a continuous measurement at 12 second interv...

Страница 20: ...ludes the combined effects of non linearity hysteresis and non repeatability as well as thermal dependencies over the compensated temperature range 3 Optional accuracy is written in compliance with US...

Страница 21: ...insure proper system operation calculate the minimum required supply voltage at the source as follows MINIMUM SUPPLY VOLTAGE 6 0 022 CABLE LENGTH x 0 07 VDC ENVIRONMENTAL Protection Rating IP68 Stora...

Страница 22: ...Revision Date Description 1 27 Sep 2018 Original 2 2 Oct 2018 Minor formatting 3 22 Oct 2018 Added use of Stage sensor extension with an Axiom SDI PT KEL used throughout 4 05 Feb Jan 2020 Removed ref...

Отзывы: