Phase
Function
1
Wait -Start with OK button-
2
Wait -can be configurated in the MENU-
4
Duration of zeroing the internal sensor
7
Duration at point of zero
9
Time of the adjustment till the next step
10
Stop time
13
Wait (target pressure)
15
Time of the adjustment till the next step
16
Duration at point of zero
18
Stop -can be configurated in the MENU-
Between the phases 10 and 16, the data -if a stable value is displayed- of the calibrator can be evaulated and recorded. Exit
of AUTO mode with the MENU button. VENT is shown left above in the display.
> LEAK (leakage)
In the CTRL mode you it is possible to check connected pressure device for leaks.
>
Connection
Pressure devices have to be connected to the + and - port of the calibrator.
Absolute pressure devices have to be connected to the + port.
>
Settings
To check pressure devices for leaks, you have to work in the CTRL mode.
>
Enter the desired pressure with the buttons UP/ DOWN or the buttons 0%/ 100%.
>
As soon as the required pressure is reached and the value is stable push the LEAK button. The internal pump will be
switched off and pressure is held.
>
Use
>
Left above: Actual mode
>
In the middle above: Target pressure
>
Right above: Unit
>
Down in the middle: Difference between actual and target pressure (in per cent)
>
Right down: Difference between actual and target pressure (absolute)
>
Left down: Time