When the calibratos displays a stable value, the data can be recorded: 

 

>

Nominal value: Right below

 

>

Unit: Right above

 

>

Actual value: On the display of the connected device

 

>

Reference value: Middle of the display

You can make a manual auto zero of the internal sensor in the CTRL mode. Connected devices can be leak tested:

 

>

Manual auto zero: Push the ZERO button - the internal sensor is get to zero.

 

>

Leak-test: Push the LEAK button to switch off the pressure cell. The pressure is held. Control the value via the display 
of the calibrator. Push the LEAK button again to stop the mode. 

> MESS (measure)

Use the MEASURE mode to measure absolute or differential pressure. The internal pump isn‘t active. The pressure is mea-
sured by the internal sensor. 

 

>

Connection

Connect differential pressure devices to the + and - port. Connect absolute pressure devices to the + port.

 

>

Setting

 

>

Push the MENU button. Confirm MODE with OK and push UP/ DOWN till MESS is displayed. Confirm with OK.

 

>

Submenu RANGE and STEPS aren‘t relevant in the MESS mode. 

 

>

Push the UP/ DOWN buttons to get to the submenu UNIT. Choose the relevant unit and confirm with the OK button. 

 

>

Push the MENU button to quit.

 

>

Submenu SETTUNGS:

 

>

Push the UP/ DOWN button to get to the item SETTINGS and confirm with OK.

 

>

Choose the item ZERO in the submenu and confirm with OK.

 

>

Via the item AUTO-ZERO you can de-activate- or activate the auto zero function.

 

>

Use

By leaving the MENU, the internal sensor goes to zero. The calibrator starts to measure the connected pressure. 

 

>

Right above: Pressure unit

 

>

Up in the middle: Maximal pressure

 

>

Left above: Mode

 

>

Middle: Actual pressure

In the MESS mode you can always make a manual zero with the ZERO button.

 

>

Attention: 

If the permissible maximum pressure (125%) is exceeded, the internal sensor is sealed off by solenoid valves. The display 
shows ERROR.  With the OK button you can unlock and release the calibrator. 

Содержание DPC

Страница 1: ...files MENU Customer specific adaption to the respecti ve application Setup Function for different operating modes Operation I O Button to switch on and off the device MENU This button will take you to the setup menu There you can adjust the settings for each mode of operation When you use the CTRL mode the calibrator according to the setup menu to leave the VENT mode LEAK This button performs a LE...

Страница 2: ...s the actual value on the LCD Connection Pressure devices have to be connected to the and port of the calibrator Absolute pressure devices have to be connected to the port Setting Press the MENU button Confirm MODE with OK and press UP DOWN until CTRL is shown on the display Confirm with OK button Set the upper limit of the range in the submenu RANGE UP DOWN and 0 100 confirm with OK Set of the un...

Страница 3: ...vices to the and port Connect absolute pressure devices to the port Setting Push the MENU button Confirm MODE with OK and push UP DOWN till MESS is displayed Confirm with OK Submenu RANGE and STEPS aren t relevant in the MESS mode Push the UP DOWN buttons to get to the submenu UNIT Choose the relevant unit and confirm with the OK button Push the MENU button to quit Submenu SETTUNGS Push the UP DOW...

Страница 4: ...Relevant pressure range UNIT Pressure unit see settings STEPS The pressure range is divided into sections Further settings of AUTO mode can be set in the submenu CONFIGURATION Cycles Set the number of cycles that have to be realized in the AUTO mode T Start Wait for vent of the internal sensor Stop Wait after reaching the maximum pressure T Pause Wait between two cycles AUTO ZERO Automatic zeroing...

Страница 5: ...NT is shown left above in the display LEAK leakage In the CTRL mode you it is possible to check connected pressure device for leaks Connection Pressure devices have to be connected to the and port of the calibrator Absolute pressure devices have to be connected to the port Settings To check pressure devices for leaks you have to work in the CTRL mode Enter the desired pressure with the buttons UP ...

Страница 6: ...owing settings can be made AKTIV If a interface is required choose of the indiviual interface USB RS232 und BAUD Choose of the correct Baudrate 1200 2400 4800 9600 14400 19200 28800 38400 56000 57600 DATA BITS Default value 8 STOP BITS Default value 1 PARITY BITS Default value N Drivers You can get the current interface drivers Windows Linux Mac OS on Homepage URL http www ftdichip com Download UR...

Отзывы: