Fault description
Cause
Remedy
The hood cannot be opened
Device fuse
Check device fuse; fuse insert on the back side
of device 2 x 10 A T (12).
Grinding stock escapes
Tensioning device (5) loose
Check and re-tension if necessary.
Seal ring is soiled or defective
Clean or replace seal ring.
Runs unevenly with strong
vibrations
Masses balanced inadequately
Set mass balance more adequately: See
Chapter 6.6 ‘Mass balance’ on page 39.
Repairs