n
DIN 66 165, part 1 basics - 1987 version
n
DIN 66 165, part 2 procedure - 1987 version
The standard sheets can be purchased from Beuth-Verlag, Berlin.
5.3 The maximum particle size permitted for test sieves
To avoid damaging the sieve base, the particle dimensions in a batch
should not exceed 10 w
0.7
(w is mesh width in mm).
Example
Nominal mesh width, w
Maximum particle size
mm
approx. mm
0.045
1
0.25
4
1.00
10
4.00
25
5.4 Recommended bulk volume for test sieves with diameters of 400 mm
Bulk volume for test sieves with a diameter
of 400 mm
Nominal mesh width
w
Feeding volume
Maximum sieve res-
idue
µm
approx. cm
3
approx. cm
3
25
44
22
32
52
26
45
60
30
63
70
34
90
84
42
125
100
50
180
120
60
250
140
70
355
160
80
Using the device