![Fresenius Medical Care T688 Series Скачать руководство пользователя страница 6](http://html.mh-extra.com/html/fresenius-medical-care/t688-series/t688-series_service-manual_2331631006.webp)
Troubleshooting
Fresenius Medical Care
T688 Series
5
V1 Aug 2020
In the event the chair fails to operate correctly, please take the following steps
prior to contacting
Fresenius Medical Care or its authorised agent.
1. Check that the mains power outlet is active.
2. Check that the power cord is not damaged or worn. Check that the chair is connected into the
mains power outlet and switched on correctly, as indicated by the green indicator light on the
nurse’s hand control.
3. Ensure that the battery has been on charge for 24 hours minimum. This should occur at least
every 5 days.
4. Ensure there are no damage or cuts in the cables to the hand controls or actuators.
5. Determine which functions are not operating as follows:
Problem / Failure
Action / Solution
The leg rest motor (vertically mounted under the
seat)
Depress the hand control button for the leg rest
and the leg rest should raise or lower.
The backrest motor (mounted behind the
backrest)
Depress the hand control button for the backrest
and the back should raise or lower.
The lift column (mounted on top of the base)
Depress the hand control button for the chair
height control and the chair should raise or
lower.
Take note of the serial number of the chair. This information can be found on the Type Label
(
Figure 1 or 2. Type Label
– Serial number location section
) attached to the chair.
If unsure about any aspect of the above steps, please contact Fresenius Medical Care Seating at the
contact details provided at the end of this service manual (see
Contact Information
section).
➢
Hand Control
– Programming
Note.This can only be done using the
nurse’s hand control. A maximum of three positions may be
programmed into the memory.
1. Position the chair to the desired position using the up & down arrows for the backrest, whole
chair and leg rest.
2. When the chair is in the desired position, hold down the S (store) button for five seconds until
the beep sounds.
3. Once the beep is sounding, select one of the numbers, 1, 2, or 3 and press it within two
seconds to store the pre-programmed position. When completed the system will beep two
times to confirm.