IFUT100-T200-T300 V5 Apr 2021  

 

 

OPERATING THE MEDICAL TREATMENT CHAIR 

Model specific features: 

T200 

– Standard chair with reclining back 

 

 

 

 

T300 

– Standard chair with reclining back & slide-out footrest 

 

 

1.  To secure the chair in position depress the brake pedals located at the top of each of 

the  rear  castors  (T200)  or  the  central  locking  brake  pedal  at  the  rear  of  the  chair 
which locks all four castors (T300).    

2.  To operate the seat/leg rest recline, release one of the lever handles on  the front side 

of the chair. To sit up, depress the legrest with your legs until the leg rest locks in the 
seated position. 

3.  To  operate  the  back  rest  (T200  &  T300  series  only);  put  the  back-release  lever 

(located  below  the  rear  arm  pad)  into  the  release  position  allowing  the  backrest  to 
move to the required position. Relock as required. 

4.  Fold out tray arm (if fitted) is accessed by opening the upper section outwards. 

5.  Lift up tray tables (if fitted) lift up and lock into place. 

6.  To allow the patient to rise from the chair, return it to the upright position. 

 

7.  Ensure procedural equipment is set up to avoid tubing and cords becoming trapped in 

the moving parts of the chair (e.g. arm rests).  DISCONNECT procedural equipment 
prior to removing patient from chair.  

For the T300 series only 

8.  The slide out footstool can be extended to allow the patient access to the chair.  The 

knob  lock  on  the  front  left  side  of  the  chair  should  be  used  to  ensure  the  stool  is  not 
allowed to slide under the chair.  The stool is normally left in the extended position for 
comfort whilst the patient is seated. 

Содержание T200 Series

Страница 1: ...Medical Treatment Chair T200 T300 Series Instruction Manual ...

Страница 2: ...5 Reporting Serious Incident 2 2 OPERATING THE MEDICAL TREATMENT CHAIR 3 3 INSTALLING THE MEDICAL TREATMENT CHAIR 4 3 1 WARNINGS AND PRECAUTIONS 4 4 MAINTENANCE 5 4 1 INSPECTION SERVICING 5 4 2 CLEANING 5 4 3 FABRIC STANDARDS 5 5 ARRANGING A SERVICE 6 6 WARRANTY INFORMATION 7 7 PHYSICAL DESCRIPTION 8 7 1 DESCRIPTION OF DEVICE 8 7 2 TECHNICAL SPECIFICATIONS 8 7 2 1 Dimensions 8 7 2 2 Operating Cond...

Страница 3: ...ted before moving the chair It is the staff members responsibility to ensure the patient is briefed on the safe operation of the chair For T200 and T300 models with CPR legs fitted In an emergency the chair can be lowered to the Trendelenburg position where CPR may be performed It is the staff members responsibility to ensure that the chair is in the correct position for CPR and that the resuscita...

Страница 4: ...T200 T300 series only put the back release lever located below the rear arm pad into the release position allowing the backrest to move to the required position Relock as required 4 Fold out tray arm if fitted is accessed by opening the upper section outwards 5 Lift up tray tables if fitted lift up and lock into place 6 To allow the patient to rise from the chair return it to the upright position ...

Страница 5: ...vers on until the chairs are in their intended operating position Note Transport damage if any should be inspected and reported immediately after delivery No claims for transport damage will be accepted 7 days after delivery date 3 1 WARNINGS AND PRECAUTIONS The following table shows potential hazards that have been identified and the steps that should be taken to avoid them IDENTIFIED HAZARD CAUS...

Страница 6: ...dure should only be used if it generally conforms to Infection Control Guidelines of the hospital Step 1 Bodily fluids may carry infectious material Ensure that protective garments and gloves are worn prior to commencing the cleaning operation Step 2 Remove as much of the spilt liquid as possible using a sponge type material with a dabbing motion Step 3 Make up a diluted bleach solution e g Milton...

Страница 7: ... serial number or your purchase order number when requesting a service The serial number can be found on the label attached to the rear of the chair An example of the label is in section 7 2 4 CONTACT INFORMATION https fmc seating com MANUFACTURER SERVICE CENTRE FRESENIUS MEDICAL CARE SEATING AUSTRALIA PTY LTD 786 Stud Road SCORESBY VICTORIA 3179 AUSTRALIA sales seatingAU fmc asia com SERVICE TEL ...

Страница 8: ...reseller this warranty covers costs at the premises of the reseller and if required in approved situations transport costs from the reseller to and from Melbourne or to some other party 4 Use of non recommended cleaning agents may void the warranty 5 The T200 T300 Series Medical Treatment Chairs are covered under a 12 month warranty 6 The T200 T300 Series Medical Treatment Chairs are not covered b...

Страница 9: ...g castors legrest reclining backrest T300 Medical Treatment Chair consists of Manually operated chair with adjustable headrest central locking castors legrest reclining backrest slide out footrest 7 2 TECHNICAL SPECIFICATIONS 7 2 1 Dimensions Model Dimensions T200 T300 Height 110 cm 115 cm Width 83 cm 83 cm Length 90 cm 90 cm Weight 41 kg 58 kg Capacity 130 kg 130 kg 7 2 2 Operating Conditions No ...

Страница 10: ...le Number Catalog Number Date of manufacture Serial Number CE mark Manufacturer For indoor use only Caution warning Consult Instruction Manual authorized representative in the European Community Medical Device Authorised Representative in the European Community Fresenius Medical Care Deutschland GmbH Else Kröner Straße 1 61352 Bad Homburg Germany Telephone 49 6172 609 0 ...

Страница 11: ...Fresenius Medical Care Seating Australia Pty Ltd 786 Stud Road Scoresby VIC 3179 Australia IFUT 100 T200 T300 EN V5 April 2021 ...

Отзывы: