Fresenius Medical Care T200 Series Скачать руководство пользователя страница 3

  FRESENIUS MEDICAL CARE       

MEDICAL TREATMENT CHAIR       

 INSTRUCTION MANUAL 

– T200 Series 

   

 

 

 T300 Series 

Page 2 of 9

 

 

 

IFUT100-T200-T300 V5 Apr 2021  

 

INTRODUCTION, PURPOSE & GENERAL SAFETY 

ISSUES 

1.1 

Indications for use 

 

 

T200 and T300 Series Medical Treatment Chairs are intended for use in medical procedures 
such  as  the  administration  of  renal  dialysis  to,  and  collecting  blood  from,  patients  in  hospital 
departments  or  home  treatment.  The  Medical  Treatment  Chairs  are  also  intended  for  use  in 
day surgery recovery areas. The chair is designed  in such a way that no special training nor 
competence is needed for operating it. 

1.2 

General Safety 

The chair should never be used in a manner for which it was not intended. 

The weight of the occupant must not exceed 130 kg. 

 
The  chair  is  designed  so  that  the  occupant  will  be  seated  in  a  typical  seated  position  -  hips 
moved back so that the occupant’s spine is against the backrest, with legs outstretched and 
supported by the seat and leg rest.  

The chair is fitted with castors, which are specifically used to aid cleaning and/or positioning of 
the chair within the room.  The chair should never be  used to transport other items.  

Ensure 

that the backrest is fully elevated before moving the chair. 

It  is  the  staff  members'  responsibility  to  ensure  the  patient  is  briefed  on  the  safe 
operation of the chair.  

For T200 and T300 models with CPR legs fitted:

 In an emergency, the chair can be lowered 

to  the  Trendelenburg  position  where  CPR  may  be  performed.  It  is  the  staff  members' 
responsibility  to  ensure  that  the  chair  is  in  the  correct  position  for  CPR  and  that  the 
resuscitation method used 

is in line with hospital policy.

 In addition, staff should ensure that 

the chair is suitable for those resuscitation methods, 

in line with hospital policy. 

In  general,  no  responsibility  or  liability  can  be  accepted  by  the  manufacturer  for  failure  to 
adhere to the guidelines and instructions contained in this manual. 

1.3 

Side Effects 

None 

1.4 

Contraindications 

There is no absolute contraindication for medical treatment chairs. But limitation of use in 
patients with history of: 

 

Allergy to covering material of chairs 

 

Pressure sores 

1.5 

Reporting Serious Incident 

Any serious incident that has occurred in relation to this medical device shall be reported to the 
manufacturer and local health authority in accordance with the local regulations. 

Within  the  EU  the  user  must  report  any  serious  incident  that  has  occurred  in  relation  to  the 

device to the manufacturer according to the labelling  

 and the competent authority of the EU 

member state in which the user is established.  

 

Содержание T200 Series

Страница 1: ...Medical Treatment Chair T200 T300 Series Instruction Manual ...

Страница 2: ...5 Reporting Serious Incident 2 2 OPERATING THE MEDICAL TREATMENT CHAIR 3 3 INSTALLING THE MEDICAL TREATMENT CHAIR 4 3 1 WARNINGS AND PRECAUTIONS 4 4 MAINTENANCE 5 4 1 INSPECTION SERVICING 5 4 2 CLEANING 5 4 3 FABRIC STANDARDS 5 5 ARRANGING A SERVICE 6 6 WARRANTY INFORMATION 7 7 PHYSICAL DESCRIPTION 8 7 1 DESCRIPTION OF DEVICE 8 7 2 TECHNICAL SPECIFICATIONS 8 7 2 1 Dimensions 8 7 2 2 Operating Cond...

Страница 3: ...ted before moving the chair It is the staff members responsibility to ensure the patient is briefed on the safe operation of the chair For T200 and T300 models with CPR legs fitted In an emergency the chair can be lowered to the Trendelenburg position where CPR may be performed It is the staff members responsibility to ensure that the chair is in the correct position for CPR and that the resuscita...

Страница 4: ...T200 T300 series only put the back release lever located below the rear arm pad into the release position allowing the backrest to move to the required position Relock as required 4 Fold out tray arm if fitted is accessed by opening the upper section outwards 5 Lift up tray tables if fitted lift up and lock into place 6 To allow the patient to rise from the chair return it to the upright position ...

Страница 5: ...vers on until the chairs are in their intended operating position Note Transport damage if any should be inspected and reported immediately after delivery No claims for transport damage will be accepted 7 days after delivery date 3 1 WARNINGS AND PRECAUTIONS The following table shows potential hazards that have been identified and the steps that should be taken to avoid them IDENTIFIED HAZARD CAUS...

Страница 6: ...dure should only be used if it generally conforms to Infection Control Guidelines of the hospital Step 1 Bodily fluids may carry infectious material Ensure that protective garments and gloves are worn prior to commencing the cleaning operation Step 2 Remove as much of the spilt liquid as possible using a sponge type material with a dabbing motion Step 3 Make up a diluted bleach solution e g Milton...

Страница 7: ... serial number or your purchase order number when requesting a service The serial number can be found on the label attached to the rear of the chair An example of the label is in section 7 2 4 CONTACT INFORMATION https fmc seating com MANUFACTURER SERVICE CENTRE FRESENIUS MEDICAL CARE SEATING AUSTRALIA PTY LTD 786 Stud Road SCORESBY VICTORIA 3179 AUSTRALIA sales seatingAU fmc asia com SERVICE TEL ...

Страница 8: ...reseller this warranty covers costs at the premises of the reseller and if required in approved situations transport costs from the reseller to and from Melbourne or to some other party 4 Use of non recommended cleaning agents may void the warranty 5 The T200 T300 Series Medical Treatment Chairs are covered under a 12 month warranty 6 The T200 T300 Series Medical Treatment Chairs are not covered b...

Страница 9: ...g castors legrest reclining backrest T300 Medical Treatment Chair consists of Manually operated chair with adjustable headrest central locking castors legrest reclining backrest slide out footrest 7 2 TECHNICAL SPECIFICATIONS 7 2 1 Dimensions Model Dimensions T200 T300 Height 110 cm 115 cm Width 83 cm 83 cm Length 90 cm 90 cm Weight 41 kg 58 kg Capacity 130 kg 130 kg 7 2 2 Operating Conditions No ...

Страница 10: ...le Number Catalog Number Date of manufacture Serial Number CE mark Manufacturer For indoor use only Caution warning Consult Instruction Manual authorized representative in the European Community Medical Device Authorised Representative in the European Community Fresenius Medical Care Deutschland GmbH Else Kröner Straße 1 61352 Bad Homburg Germany Telephone 49 6172 609 0 ...

Страница 11: ...Fresenius Medical Care Seating Australia Pty Ltd 786 Stud Road Scoresby VIC 3179 Australia IFUT 100 T200 T300 EN V5 April 2021 ...

Отзывы: