timer indicating how much time is left to finish the test. A “Please Wait”
message will be flashing on and off on line 2 during this test.
F1 F2 F3 F4
Upon test completion, the final CAL-V™ value will be displayed along with a
Pass/Fail message.
Performing the Zero CAL-CHECK™ Calibration Validation Test
The Zero CAL-CHECK™ test is a companion test to CAL-V™. Unlike CAL-V™,
which may be performed in the pipe and at process conditions, Zero CAL-
CHECK™ must be performed at zero flow to ensure a valid test result. This
test is used to confirm that the flow meter still retains its original NIST-
traceable calibration at zero flow and that the sensor is free of film or residue
tat may affect readings. The test takes less than 5 minutes to complete. At the
conclusion of the test, a Pass or Fail message will be displayed. Press
F4
at
the conclusion of the test to return to normal measuring mode or to terminate
the test.
Press
FLO (F2)
from the main menu. The display will show:
F1 F2 F3 F4
Press
DGN (F1).
The display will show:
F1 F2 F3 F4
Press
TST (F2).
The display will show:
FLOW PARAMETER 1
DGN UNT PRM EXIT
DIAGNOSTIC
SIM TST EXIT
18
Calibration Validation
OPERA
TION
Operation: Zero CAL-CHECK
™
CAL-V
™
Model FT3
CAL-V = 0.51
Passed OK
Zero CAL-CHECK
™