![fos4X Blackbird OECD Скачать руководство пользователя страница 45](http://html1.mh-extra.com/html/fos4x/blackbird-oecd/blackbird-oecd_instructions-for-use-manual_2322762045.webp)
3. Product Description
Table 38: Identity object
Name
Content
Vendor-ID
0x0000f4f4
Product code
MDE1261
Revision number
Software Version (major version in upper 16 bit,
minor version in lower 16 bit)
Serial number
32 bit serial number of the measuring device
0x2000 Wavelength Data
Description
All values are 16 bit signed numbers.
Values between -32752 (0x800F) and -32768 (0x8000) are error values. See
Error values in 0x2000, page 45
The other values are formed with the offset and scaling factor for each channel. See
and
Offset and scaling factor can be configured via the web-interface. See
.
Error Values
Table 39: Error values in 0x2000
Value
Description
0x8000
Other sensor error.
0x8001
Channel not available on this measurement
device.
0x8002
No sensor detected on this channel.
0x8003
Sensor temporarily unavailable for measurement
(light source regulates power).
0x8004
Sensor or channel error.
0x8005 … 0x800F
Reserved.
0x2001 Status
The Blackbird OECD has 4 status fields.
© fos4X 2020
MDE1261_4-10_1.3.x_5.0_Blackbird
45 / 83
Содержание Blackbird OECD
Страница 83: ......