![formlabs Form Wash L Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/formlabs/form-wash-l/form-wash-l_manual_2320172034.webp)
32
8. Dispose of used solvent according to local regulations.
6.4.2
Cleaning the wash bucket
Over time, dissolved resin accumulates on the walls of the wash bucket. Periodically, while
replacing the solvent, clean out the wash bucket to remove these resin deposits. Rinse the
wash bucket with fresh solvent and wipe with a paper towel.
Required supplies:
•
2.5 mm hex driver
•
43 L (11.4 US gal) of fresh solvent (IPA, TPM, or water)
•
Electric siphon pump, included with the Form Wash L
•
Empty container, suitable for storing up to 37.9 L (10 US gal) of solvent
•
Non-reactive nitrile gloves
•
Safety glasses
Follow these steps to transfer solvent from the Form Wash L into a chemically-resistant
container that is safely designated for highly-resin-concentrated solvent and to refill the wash
bucket with fresh solvent.
To clean the wash bucket:
1. Empty the wash bucket and remove it from the Form Wash L. Refer to section
6.4.1
Replacing solvent
for detailed instructions.
2. There are two circular impellers located at the bottom of the wash bucket. Using a 2.5 mm
hex driver, loosen and remove the screw holding each impeller in place. Set the screws
and impellers aside.
3. Inspect the bearing underneath the impellers and the bottom surface of the impellers for signs of
wear. Visible wear on either surface is a sign that the impeller and bearing should be replaced.
4. Rinse the wash bucket and solvent monitor with IPA or water and wipe with a paper towel.
5. Reseat the two impellers in the wash bucket. Thread one screw into each of the impellers.
Using a 2.5 mm hex driver, tighten the impeller screws until finger tight.
6. Reinstall the wash bucket and fill it with solvent. Refer to section
6.4.1 Replacing solvent
for detailed instructions.
6.5
Intermittent maintenance
Task
Frequency
Refer to
Section
Update the firmware
When indicated by Formlabs
Updating firmware
4.7
Dispose of used solvent
After replacing solvent
Disposal of solvent 9.3
Replace the wash bucket
After 2,000 hours of use
Replacing the
wash bucket
6.5.1
6.5.1
Replacing the wash bucket
The wash bucket is rated for up to 2,000 hours of use. After this length of time, the bucket
should be replaced.
To replace the wash bucket:
1. Lift the Form Wash L lid.
2. Using the electric siphon pump, pump all solvent out of the wash bucket and into a container
appropriate for your solvent of choice.
a.
Place the pump intake into the wash bucket.
Содержание Form Wash L