formlabs Form Wash L Скачать руководство пользователя страница 29

Usage | 

27

5.4 

Time settings

For detailed guidance and visual assistance, visit 

support.formlabs.com

.

For the best results, use the recommended time settings tested specifically for use with the 
Form Wash L.
Biocompatible materials are a speciality product, developed for use by medical professionals, 
and should be used in accordance with the instructions for use.

5.5 

Managing the machine

Between wash cycles, perform maintenance, manage your solvent usage, track the machine’s 
history, or turn the machine off. For detailed guidance and visual assistance, visit 

support.

formlabs.com

.

5.5.1 

Extending solvent lifetime

The solvent in the Form Wash L builds up a higher concentration of resin after each wash, 
as liquid resin from each wash cycle accumulates over time. As the solvent becomes 
more resin-concentrated, a layer of diluted resin coats the outermost surface of printed 
parts after washing. As the solvent dries, a thin layer of liquid resin is left behind on the 
part’s surface, causing the surface to feel tacky. Thus, the part is only as clean as the 
cleaning solution.
When washing parts on the build platform, scrape excess liquid resin back into the resin tank 
before placing the build platform in the Form Wash L. To achieve the cleanest parts, replace 
used solvent with fresh solvent frequently or consider a multi-step washing process. As solvent 
evaporates, add fresh solvent to keep the fill level between the minimum and maximum lines.
Maintain separate wash buckets for each resin type or similar colors to preserve the best 
possible surface quality and performance. For example:

 •

Biocompatible resins – to comply with biocompatibility regulations

 •

Castable Wax 40 Resin – to avoid color transfer

 •

Castable Wax Resin – to avoid color transfer

 •

Rigid 4000 Resin – to prevent glass particles adhering to parts printed with other resins

 •

Rigid 10K Resin – to prevent glass particles adhering to parts printed with other resins

 •

Ceramic Resin – to prevent ceramic particles adhering to parts printed with other resins

Solvent lifetime depends in part on the chemical properties of the solvent. Visit 

support.

formlabs.com

 for detailed information about different solvents and solvent performance.

5.5.2 

Measuring the resin concentration of solvent

As more parts are cleaned, the solvent gradually becomes more concentrated with liquid 
resin and reduces the effectiveness of the Form Wash L . Replace solvent when the wash is 
no longer effective and when parts have tacky surfaces after washing.
To significantly prolong the lifetime of the solvent before replacement, consider implementing 
a preliminary rinse with a small volume of solvent before inserting printed parts into the Form 
Wash L.

To measure the resin concentration of solvent with the solvent monitor:

1.  If you are currently viewing a submenu on the touchscreen, tap the < arrow until you return 

to the 

Home

 screen. The home screen lists your current wash settings and the status of 

the Form Wash L.

2.  The current solvent type and resin concentration are listed under 

Solvent

.

Содержание Form Wash L

Страница 1: ...Manual Form Wash L...

Страница 2: ......

Страница 3: ...and Usage Instructions Form Wash L Large format Desktop Stereolithography SLA Print Cleaner Original English instructions Read this manual carefully and keep it for future reference November 2021 REV...

Страница 4: ...14 3 4 Sensitive components 15 3 5 Emergency and exceptional situations 15 4 Preparation and setup 17 4 1 Location and environs 17 4 2 Power and networking 17 4 3 Unboxing the machine 17 4 4 Accessin...

Страница 5: ...eset 34 7 3 Cleaning after a chemical spill 34 7 4 Troubleshooting errors or abnormal activity 35 8 Disassembly and repair 38 8 1 Tasks 38 9 Recycling and disposal 39 9 1 Disposal of resin 39 9 2 Recy...

Страница 6: ......

Страница 7: ...or integrator to perform the appropriate and complete risk analysis evaluation and testing of the products with respect to the relevant specific application or use thereof Neither Formlabs nor any of...

Страница 8: ...njuries The Form Wash L shall only be used by persons who have fully read and understand the contents of this manual Ensure that each person who uses the Form Wash L has read these warnings and instru...

Страница 9: ...mlabs Support support formlabs com USA Formlabs Inc 35 Medford St Somerville MA USA 02143 Germany Formlabs GmbH Nalepastrasse 18 12459 Berlin Germany 1 2 2 Warranty This product is protected under war...

Страница 10: ...SSLY DISCLAIMS ANY IMPLIED OR EXPLICIT WARRANTY OF FITNESS for a particular usage the particular nature and circumstances of said usage being unforeseen and unforeseeable to Formlabs Formlabs is not a...

Страница 11: ...submerge solvent monitor Maximum part size W D H 33 5 20 30 cm 13 2 7 9 11 8 in Compatible solvents Isopropyl alcohol IPA Tripropylene glycol monomethyl ether TPM Water Operating environment 18 28 C 6...

Страница 12: ...be lowered or raised from the bucket Wash bucket Removable container holds a maximum of 43 L 11 4 US gal of solvent A rotating impeller at the bottom circulates the solvent Solvent monitor A sensor me...

Страница 13: ...ump A battery powered siphon pump for moving solvent into and out of the wash bucket The pump has a dedicated storage slot underneath the wash bucket 2 4 User interface For detailed guidance and visua...

Страница 14: ...N indicates a hazard with a low level of risk which if not avoided could result in minor or moderate injury NOTICE indicates information considered important but not hazard related DANGER Isopropyl al...

Страница 15: ...l applications on the human body Refer to the safety data sheet SDS for each specific resin as well as support formlabs com for more detail 3 1 3 Radio interference This equipment has been tested and...

Страница 16: ...fety information Carefully follow the safety instructions provided with the tripropylene glycol monomethyl ether you purchase We also recommend that you wear protective gloves when working with tripro...

Страница 17: ...st government guidelines Always consult the safety data sheet SDS as the primary source of information to understand safety and handling of Formlabs materials and required accessories Visit formlabs c...

Страница 18: ...sin spill to minimize any cosmetic or functional damage to the machine If you have experienced an accidental resin spill document the problem with photos and clean the device as best as possible Conta...

Страница 19: ...Power and networking For detailed guidance and visual assistance visit support formlabs com The Form Wash L requires reliable grounding The power supply used with the device must be grounded Do not us...

Страница 20: ...le via the touchscreen To access the serial name on the touchscreen 1 If you are currently viewing a submenu on the touchscreen tap the arrow until you return to the home screen The home screen lists...

Страница 21: ...that is required to connect over Wi Fi or Ethernet Visit support apple com for assistance with Bonjour The USB connection can still be used while the device is connected to a LAN When the device is c...

Страница 22: ...irmware file on the machine Refer to the firmware downloads and release notes to learn more about the improvements that come with each version s release To update the firmware 1 Open PreForm 2 Connect...

Страница 23: ...ch may result in additional fees void the warranty or present a safety hazard Do not ship the Form Wash L with any loose items stored inside the machine Loose items can shift during transit and damage...

Страница 24: ...lift the machine into its carrier tray Lifting hazard The Form Wash L is a heavy object A two person lift is required to prevent injury and avoid damage to the machine 5 Place the foam insert on the t...

Страница 25: ...efore or after they have been separated from the build platform To insert printed parts for cleaning 1 Use the touchscreen to raise the platform and basket mounts 2 Insert printed parts a Form 3L Buil...

Страница 26: ...the parts you are washing tap Quick Wash Default Wash or Long Wash for a 5 10 or 20 minute wash cycle respectively Tap the i icon under each preset to learn more about the preset or tap Guidelines to...

Страница 27: ...ile the solvent evaporates Refer to section 5 1 Operational environment for more information Observe necessary safety precautions according to the solvent supplier s SDS Options for drying include air...

Страница 28: ...solvent drains out of the chamber before transferring the part to another surface 5 3 2 Internal channels Thin channels such as those used in micro fluidic designs may not drain entirely on their own...

Страница 29: ...n the minimum and maximum lines Maintain separate wash buckets for each resin type or similar colors to preserve the best possible surface quality and performance For example Biocompatible resins to c...

Страница 30: ...to Dashboard through the touchscreen and use the reported metrics to plan or schedule your usage of the Form Wash L 5 5 4 Turning off The Form Wash L is designed to remain powered on when not in use F...

Страница 31: ...ed When removing the exterior paneling disconnect the machine from its power source before maintenance Moving parts present crushing and tangling hazards Formlabs provides instructions to advise skill...

Страница 32: ...n the minimum and maximum fill lines If the solvent level is below the minimum fill line parts washed on the build platform may not be fully immersed in the solvent during the wash To add solvent to t...

Страница 33: ...nt rises out of the wash bucket b Open the Form Wash L door Gently pull the wash bucket towards yourself to remove it from the Form Wash L The wash bucket is still connected to the machine via the sol...

Страница 34: ...ach impeller in place Set the screws and impellers aside 3 Inspect the bearing underneath the impellers and the bottom surface of the impellers for signs of wear Visible wear on either surface is a si...

Страница 35: ...me so place the wash bucket next to the Form Wash L c Lift the inner lid off of the wash bucket Set it aside d Unclip the solvent monitor bracket from the front edge of the wash bucket Lift the solven...

Страница 36: ...ency tools in a known accessible location When working with isopropyl alcohol or other flammable chemicals consider purchasing a flammable solvent cleanup kit Required supplies Apron Non reactive nitr...

Страница 37: ...ons Error Cause Solution The display does not turn on Power failure or a faulty electrical connection Disconnect and reconnect the power Plug the power cable into a different outlet The touchscreen is...

Страница 38: ...o raise the basket mount and platform mount when they were already at their top position Disconnect and reconnect the power The device position is unknown The Form Wash L attempted to raise the basket...

Страница 39: ...wn error occurred Other unspecified cause Disconnect and reconnect the power Any other error related to the solvent monitor The Form Wash L detected an issue with the solvent monitor Check that the so...

Страница 40: ...om its power source before maintenance 8 1 Tasks Contact Formlabs Support or a certified service provider to receive repair instructions and authorization including how to disassemble or remove the ex...

Страница 41: ...nmlightand heat for the most effective curing The liquid resin cures to solid material when exposed to light and heat 4 Dispose of the fully cured resin and container as household waste 9 1 2 Cured re...

Страница 42: ...essories or packaging indicates that this device shall not be treated as nor disposed of with household waste When you decide to dispose of this product do so in accordance with environmental laws and...

Страница 43: ...nce 5 12 27 29 30 32 38 41 motor 35 36 41 42 43 O operation 6 12 14 41 P power 11 13 15 17 18 19 21 23 28 29 34 35 36 37 38 41 43 R resin 5 6 8 10 11 12 13 14 15 16 23 24 25 26 27 30 31 32 39 40 41 43...

Страница 44: ...unt and basket mount raise and lower Motherboard The motherboard is the main circuitry through which all systems in the Form Wash L communicate Outer lid Limits solvent evaporation Keep the lid closed...

Страница 45: ...ith the following electronics and safety standards ETL CE UL 61010 1 CAN CSA C22 2 No 61010 1 12 Machinery Directive 2006 42 EC Radio Equipment Directive 2014 53 EU RoHS Directive 2011 65 EU EMC Direc...

Страница 46: ...Original English instructions November 2021 Formlabs...

Отзывы: