background image

 

2. Assembling and flow check the Limited-Use Public Health Emergency PAPR system 

 

Attach blower unit to belt (the filter comes pre-installed in the blower unit) 

 

Open battery compartment cover, releasing both latching tabs, one at a time.   

 

Attach fully charged battery pack to adapter, do not use excessive force when extracting adapter  

 

Place battery/adapter inside of the blower unit with foam block and close & latch battery compartment cover 

 

Perform flow check (prior to each use) 

-

 

Ensure float ball in the Air Flow Indicator moves freely 

-

 

Insert Air Flow Indicator into the outlet of the blower unit 

-

 

Start blower unit and run for 1 minute 

-

 

With the Air Flow Indicator in a vertical position, the bottom of the floating ball must be at, or above, the minimum flow mark 

-

 

If float ball fails to rise above the minimum mark, do not use the blower unit 

-

 

Remove Air Flow Indicator from the blower unit 

 

Attach the “TO BLOWER” end of the breathing tube onto the blower unit air outlet 

 

Gently pull on breathing tube to ensure secure connection to blower unit 

 

Connect the breathing tube to the hoodtop by pushing the breathing tube onto the quick connector until it snaps into place 

 

Gently pull on breathing tube to ensure secure connection to hoodtop 

3. Donning the Limited-Use Public Health Emergency PAPR system 

 

Use belt to attach the PAPR to wearer’s waist securely 

 

Pull hoodtop over head, adjusting face seal under chin making sure facial hair does not interfere with the face seal 

4. Doffing and post-use of the Limited-Use Public Health Emergency PAPR system 

 

Exit the contaminated work area 

 

Remove the PAPR system making sure to follow workplace procedures so as not to expose yourself to contaminants 

 

Turn off blower unit by pressing the power button on the bottom of the blower unit 

 

Clean and store the PAPR following practices established by your employer and the user instructions 

 

 

 

Ford Motor Company 
One American Road  
Dearborn, MI  48126 

Occupational Use Only 
 

In United States of America 
Technical Service: 833-804-3673 

Customer Service: 833-804-3673 
www.corporate.ford.com/ppe

 

 

Please recycle. 

Отзывы: