Section 6
Commissioning
Page: 32
CODEL
A Forbes Marshall Company
Issue: A | Revision: 9 | Date: 15/10/21 | Reference: 100246
1.19
Analogue (mA) Output Configuration:
The mA output menu allows the user to configure the measurand unit, zero and
span information for the device’s mA Outputs. In addition, the menu allows the
user to simulate a given value on any attached board to test each individual
channel.
NOTE: the values and menu choices below are for example only, and/or are default
values. When completing the commissioning process, fill these fields to the user’s
specification.
The ma Output unit - Sets what unit or measurement is used for the mA output.
Enter the menu path:
Setup Mode → mA outputs → mA Output unit
•
Set the mA Output Unit to the customer specification and press enter to open the
Set Output Unit menu option.
Main Menu
Submenu 1
Submenu 2
Setup Mode
mA Outputs
mA Output Unit
Span
Zero
mA Simulation
Exit