background image

3. Apply Control Solution 

Shake the control solution vial thoroughly before 

use. 

Squeeze out a drop and wipe it off, then squeeze 

out another drop and place it on the tip of the vial 

cap. Hold the meter to move the absorbent hole 

of the test strip to touch the drop. Once the 

confirmation window fills completely, the meter 

will begin counting down. To avoid contaminating 

the control solution, do not directly apply control 

solution onto a strip.

109~165

Normal

When “QC” is selected, the control solution test 

result will appear on the screen however the 

meter will not store the test result in the memory 

of iPhone.

WARNING:

When doing the control solution test, you have to 

mark it so that the test result will not be stored in 

the memory. Failure to do so will mix up the 

blood glucose test results with the control 

solution test results in memory.

2. Select the QC option on the screen to mark 

this test as s control solution test.

● The meter provides you with five measurement 

modes, each are:

4. Read and Compare the Result 

After “Processing” message appears, the control 

solution test result will then show on the display. 

Compare this result with the range printed on the 

test strip vial and it should fall within this range. If 

not, please read the instructions again and 

repeat the control solution test.

An example: (117mg/dL = 6.5 mmol/L; 109-165 

mg/dL =6.1-9.2 mmol/L)

Press the lancing device’s tip firmly against the 

lower side of your fingertip. Press the release 

button to prick your finger, then a click indicates 

that the puncture is complete.

Please follow the suggestions below before 

obtaining a drop of blood:

● Wash and dry your hands before starting.

● Select the puncture site either at fingertips or 

another body parts (please see section 

“Obtaining a Blood Sample from Alternative 

Sites” (AST) on how to select the appropriate 

sites). 

● Clean the puncture site using cotton 

moistened with 70% alcohol and let it air dry. 

● Rub the puncture site for about 20 seconds 

before penetration.

● Use a clear cap (included in the kit) while 

setting up the lancing device.

Out-of-range results

If you continue to have test results fall outside 

the range printed on the test strip vial, the meter 

and strips may not be working properly. Do NOT 

test your blood. Contact the local customer 

service or place of purchase for help.

NOTE:

● The control solution range printed on the test 

strip vial is for control solution use only. It is not 

a recommended range for your blood glucose 

level.

● See the Caring for Your iDiamond Meter Set 

section for important information about your 

control solutions.

Our Control Solution contains a known amount of 

glucose that reacts with test strips and is used to 

ensure your meter and test strips are working 

together correctly. 

Test strips, control solutions, or sterile lancets 

may not be included in the kit (please check the 

contents on your product box). They can be 

purchased separately. Please make sure you 

have those items needed for a blood glucose test 

beforehand.

Do a control solution test when:

● you first receive the meter,

● at least once a week to routinely check the 

meter and test strips,

● you begin using a new vial of test strips,

● you suspect the meter or test strips are not 

working properly,

● your blood glucose test results are not 

consistent with how you feel, or if you think the 

results are not accurate,

● practicing the testing process, or

● you have dropped or think you may have 

damaged the meter.

Important Information about 

Performing Control Solution Tests

You can select the required mode by:

selecting between AM, AC, PC, PM and QC 

mode.

AM

QC

PM

AC

PC

Data

Plan

FAQ’s

Settings

Take blood samples

QC

QC

2011/12/25                        AM 09:15

mmol/L

117

Fingertip testing

Blood from sites other than the fingertip

Intended Use

Test Principle

This system is intended for use outside the body 

(in vitro diagnostic use) by people with diabetes 

at home and by health care professionals in 

clinical setting as an aid to monitoring the 

effectiveness of diabetes control. It is intended to 

be used for the quantitative measurement of 

glucose (sugar) in fresh whole blood samples 

(from the finger, palm, forearm, upper arm, calf 

and thigh). It should not be used for the 

diagnosis of diabetes, or testing on newborns.

Professionals may test with capillary and venous 

blood sample; home use is limited to capillary 

whole blood testing. Venous blood must be 

collected only in heparin blood collection tube.

Your system measures the amount of sugar 

(glucose) in whole blood. The glucose testing is 

based on the measurement of electrical current 

generated by the reaction of glucose with the 

reagent of the strip. The meter measures the 

current, calculates the blood glucose level, and 

displays the result. The strength of the current 

produced by the reaction depends on the amount 

of glucose in the blood sample.

How to Perform a Control Solution Test

1. Plug in iDiamond

Connect iDiamond with iPhone, a “connecting” 

message will appear on the screen and follows 

by the message “insert the test strip into the 

meter”. Please follow the instruction and insert 

the test strip into iDiamond.

You will need:

1. Meter

2. Test Strip

3. Control Solution

Data

Plan

FAQ’s

Settings

Insert the test strip 

into the meter

Connecting

WARNING:

To reduce the chance of infection: 

● Never share a lancet or the lancing device. 

● Always use a new, sterile lancet. Lancets are 

for single use only. 

● Avoid getting hand lotion, oils, dirt, or debris in 

or on the lancets and the lancing device.

Preparing the Lancing Device for Blood 

Testing

Please follow the instructions in the lancing 

device insert for collecting a blood sample.

Preparing the Puncture Site

Stimulate blood perfusion by rubbing the 

puncture site before blood extraction has a 

significant influence on the glucose value 

obtained. 

Blood from a site that has not been rubbed 

exhibits a measurably different glucose 

concentration than blood from the finger. When 

the puncture site was rubbed prior to blood 

extraction, the difference was significantly 

reduced.

Important Information about 

Testing Your Blood Glucose Level

Replace the lancing device cap with the clear cap 

for AST. Pull the cocking control back until it 

clicks. When lancing the forearm, upper arm, 

hand, thigh, or calf, avoid lancing the areas with 

obvious veins because of excessive bleeding. 

NOTE:

● Choose a different spot each time you test. 

Repeated punctures at the same spot may 

cause soreness and calluses.

● Please consult your health care professional 

before you begin AST.

● It is recommended that you discard the first 

drop of blood as it might contain tissue fluid, 

which may affect the test result.

QC

Содержание iDiamond

Страница 1: ...des important information to help you to use the system properly Before using this product please read the following contents thoroughly and carefully Regular monitoring of your blood glucose levels c...

Страница 2: ...e not working properly your blood glucose test results are not consistent with how you feel or if you think the results are not accurate practicing the testing process or you have dropped or think you...

Страница 3: ...ycemia Your AST results do not match the way you feel Your routine glucose results often fluctuate AM PM AC PC QC QC QC wake up in the morning before meal about 2 hours after a meal in the late evenin...

Страница 4: ...lect the enlarge button on the top right corner to switch between 7 14 30days view To see different times simply slide to the left or right Swipe up to see the next function Pie Chart By swiping to th...

Страница 5: ...owledge their monitoring condition over a certain period of time Reviewing Target Pie Chart PLAN Reviewing Logbook Reviewing Trend Graph Data Plan FAQ s Settings 2011 12 20 18 30 97 mg dL 2011 12 21 1...

Страница 6: ...tion to examine whether those daily action can achieve your goal to control glucose WARNING While the meter is uploading data to the TeleHealth it will be unable to perform a blood glucose test Error...

Страница 7: ...o the test strip package insert Important Control Solution Information Use only our control solutions with your meter Do not use the control solution beyond the expiry date or 3 months after first ope...

Отзывы: