background image

Where can I review my data?

Since the D40g is not Bluetooth-enabled, it will not connect to an 
app on a smartphone. You can check the memory on the D40g for 
the latest 864 readings by pressing the M button or log in to the 
24/7 HealthView Telehealth patient portal at telehealth.foracare.
com to review your latest readings and keep track of  your trends. 
The product user manual can be downloaded from the website 
product page.

What is control solution and how do I use it?

Control solution is used to check if  the meter and test strips are 
properly working together as a system. You only need control 
solution if  you are using the meter to measure your blood glucose 
levels, and only when you get a new set of  test strips or think your 
results are not accurate. To use the control solution, insert the test 
strip into the meter as usual, apply the second drop of  control 
solution to the test strip (do not apply control solution directly on 
to the test strip), and compare the results with the range printed 
on the control solution vial. Please note the expiration date, and 
discard any control solution that has expired or been open for 
longer than 3 months.

What are the cleaning and disinfecting procedures for the D40g?

Clean the meter when you see any dirt on it. Disinfect the meter 
at least once a week to prevent infection. To clean, use one 
disinfecting wipe on exposed surfaces of  the meter to remove 
visible dirt, blood or other fluids. Do not use organic solvents to 
clean the meter. To disinfect, clean the meter first. Then grab one 
disinfecting wipe and squeeze out any excess liquid to prevent 
damage. Wipe all meter’s exterior surface display and buttons, 
but be careful not to allow excess liquid to get inside the test slot. 
Keep meter wet with disinfecting solution for a minimum of  two 
minutes. Remove the wipe and allow the surface to dry completely. 
Discard the wipe. Wash your hands thoroughly after handling the 
meter, lancing device, and test strips to avoid contamination. This 
device has been validated to withstand up to 6,000 cleaning and 
disinfecting cycles using the recommended disinfecting wipes. 
See the user manual for more information.

For additional information on the D40g, please refer to the user 
manual.

FAQ

How does the D40g work?

The D40g measures the systolic and diastolic blood pressure 
and pulse rate by using a non-invasive technique in which an 
inflatable cuff  is wrapped on the upper arm. The device also 
measures blood glucose from fresh capillary whole blood 
samples. It automatically uploads data to the 24/7 HealthView 
Telehealth System using 3G cellular technology. 

What’s the proper way to take a measurement?

To test your blood pressure, make sure you’ve been sitting for at 
least 10 minutes, your elbow is on a flat surface with your palm 
facing up, and the cuff  is about the same level as your heart. Do 
not talk or move during measurement. To test your blood glucose 
levels, wash and dry your hands before starting, insert the test 
strip into the device, rub the puncture site for about 20 seconds, 
puncture your finger, wipe away the first drop of  blood, squeeze 
the finger to produce a larger sample size, and carefully touch 
the finger to the tip of  the test strip allowing the absorbent hole to 
sip in the blood.

How often do I need to charge the battery?

Please charge battery for 5 hours before first use. The lifespan 
of  your battery will vary considerably with how it is used, how it 
is maintained and charged, temperature, and other factors such 
as frequency of  testing. Use the A/C adaptor (input: 100-240V / 
output: 6V 1A) to recharge the 3.7V rechargeable Li-ion battery. 
After about 5 hours, the battery will be fully charged, as indicated 
when the green battery light disappears.

Can I use the meter without the battery?

Yes, you can use the device to measure your blood pressure 
when it is plugged in to the A/C adaptor without a battery. 
However, it will not upload any readings until you insert the 
battery again. You will not lose any readings while the battery is 
not inserted.

Содержание D40g

Страница 1: ...D40g Blood Pressure and Blood Glucose Monitoring System Quick Start Guide EN ...

Страница 2: ...0g Meter Information Intended for single patient use only Meter will switch off automatically if left idle for 3 minutes Meter should transmit data automatically For more information on manual data transmission or signal quality check please refer to the user manual Blood Glucose Strips Strips have an expiration date Please check strip container for the expiration date before using Strips expire 9...

Страница 3: ...wer button All the LCD symbols will appear Then the cuff will begin to inflate automatically 7 The heart symbol will flash when a pulse is detected during the inflation 8 After the measurement the monitor displays the systolic pressure diastolic pressure and pulse rate Performing a Test Blood Glucose 1 Wash your hands 2 Insert lancet into FORA lancing device and close 3 Insert test strip into the ...

Страница 4: ...ter handling the meter lancing device and test strips to avoid contamination This device has been validated to withstand up to 6 000 cleaning and disinfecting cycles using the recommended disinfecting wipes See the user manual for more information For additional information on the D40g please refer to the user manual FAQ How does the D40g work The D40g measures the systolic and diastolic blood pre...

Страница 5: ...ine at 1 888 307 8188 Error Code Err 1 Err 2 Appears when the batteries cannot provide enough power for a test Replace or recharge the batteries immediately PCL Err 30 Unable to connect to the service PCL Err 40 PCL Err 41 PCL Err 42 PCL Err 43 PCL Err 44 PCL Err 45 PCL Err 48 PCL Err 49 No response from the server What It Means Solution Err 3 Err 4 Err 5 No pulse rate detected or the pulse rate i...