Blood Glucose Measurement Performance
Measurement Unit:
mg/dL
Measurement Range:
20 to 600 mg/dL (1.1 to 33.3 mmol/L)
Precision:
± 5 % (CV)
Accuracy:
±15 mg/dL (0.83 mmol/L) when glucose < 75mg/dL (4.16 mmol/L)
± 20% when glucose
≥
75 mg/dL (4.16 mmol/L)
Ketone Warning:
≥
240 mg/dL (13.3 mmol/L)
Blood Pressure Measurement Performance
Pressure Range:
0 to 300 mmHg
Measurement Unit:
mmHg
Systolic Measurement Range:
60 to 250 mmHg
Diastolic Measurement Range:
30 to 180 mmHg
Pulse Rate Measurement Range:
40 to 199 beats / minute
Maximum Inflation Pressure:
280 mmHg
Accuracy of Pressure:
±3 mmHg or ±2% of reading
Accuracy of Pulse Rate:
±4% of reading
This device has been tested to meet the electrical and safety requirements of: IEC/ EN
60601-1, IEC/EN 60601-1-2, IEC/EN 61010-1, IEC/EN 61010-2-101, IEC/EN
61326-2-6.
Reference to Standards:
• IEC 60601-1 General requirement for safety
• IEC 60601-1-2 Requirements for EMC
• AAMI/ANSI /IEC 80601-2-30, ANSI/AAMI/ISO 81060-2, NIBP requirements