39
MESSAGE
WHAT IT MEANS
WHAT TO DO
Err 1
Err 2
Inflation or pressure error.
Refit cuff tightly and
correctly. Relax and repeat
the measurement. If the
error still remains, please
contact local customer
service for help.
Err 3
Cuff pressure is too high.
Err 4
Err 5
No pulse rate detected or
the pulse rate is too low.
Err 6
Systolic pressure is outside
the system’s measurement
range
System measurement range
is 60 mmHg – 250 mmHg.
Review the instructions and
repeat the test. If the error
still remains, please contact
your healthcare professional
for help.
Err 8
Forced to stop the blood
pressure measurement after
re-pump for 2 times.
Turn off the device and start
over. Follow the instructions
and try again.
Appears when the batteries
cannot provide enough
power for a test.
Replace or recharge the
batteries immediately.
E-U
Strip has been used.
Repeat the test with a new
test strip. If the meter still
does not work, please
contact the customer service
for assistance.
E-t HIGH
E-t LOW
Environmental temperature
is outside the system’s
operational range.
System operational range is
50°F to 104°F (10°C to
40°C). Repeat the test after
the meter and test strip have
reached the above
temperature.