EN-5
Note:
• Press to select Beep On, Universal Tone On or Beep Off.
Press MAIN button to confirm.
• When Universal Tone is turned on, the meter guides you
through the blood glucose test using beep tones; it also
outputs the result as a series of beeps.
• When Beep is turned off, the alarm function will remain
effective.
• During memory deletion, select “ ”to keep all saved
results.
• You may set it up to four reminder alarms.
• To turn off the alarm, press or to change On to OFF.
Press
MAIN
button to confirm.
• When the alarm goes off, the device will automatically turn
on. Press to mute the alarm. If you do not press , the
device will beep for 2 minutes then switch off.
• If the device is idle for 3 minutes during the setting mode,
it will turn off automatically.
Before Testing
Calibration
You must calibrate the meter every time you begin to use
a new vial of β-ketone test strips by setting the meter with
the correct code. Test results may be inaccurate if the code
number displayed on the monitor does not match the
number printed on the strip label or strip foil pack.
How to Code Your Meter (for β-ketone test)
1. Insert the code strip when the monitor is off. Wait until the
code number appears on the display.
Note:
Make sure the code numbers on display, code strip, and test
strip vial or foil pack are the same. The code strip should be
within the expiry date; otherwise, an error message may
appear.
2. Remove the code strip, the display will show “OFF”. This
tells you that the meter has finished coding and is ready for
β-ketone testing.
Checking the Code Number
You need to make sure that the code number displayed on
the meter matches the number on your test strip vial or foil
pack before you proceed. If it matches, you can proceed with