4.5 Switching on motor
OPERATION
40
6.452
4.5 Switching
on
motor
NOTE
!
Check the diameter and height settings of the sensor system before
switching on.
The operating elements used to switch on the rotating head drive are lo-
cated on the motor control.
The motor control is installed in a separate housing.
Acceleration times:
approx. 4 sec (3,000) rpm
approx. 10 sec (6,000) rpm
Deceleration times without braking: approx. 210 sec (3,000) rpm
approx. 360 sec (6,000) rpm
Deceleration times with braking:
approx. 15 sec (3,000) rpm
approx. 30 sec (6,000) rpm
The rotating head must run quietly and without loud noises when no test
material is present.
Whistling noises when starting or braking indicate that the V-ribbed belt
needs to be retensioned.
NOTE
!
Switch off the drive immediately if grinding noises are heard.
Possible causes:
Incorrect nozzles inserted
Nozzle or holders not securely fitted
Nozzle not fitted in end position