Frequently asked questions
Unit does not respond
Check if the unit is online. If not, then check the router handles 2.4 GHZ, and if a
firewall in the router is blocking the communication.
LCD says that leakage has been detected
Mastertronic has an inbuilt leakage sensor for security reasons. If something
unexpected happens and leakage occurs, the machine will stop immediately and you
shall receive a notification. You will have to dry the machine, and then press
”Leakage resolved ” and Mastertronic will resume.
For drying properly do this:
1) Dry up the visible water that is on the leakage sensor.
2) Press ”
Leakage Resolved
” and if sensor is dry, it
will resume.
3) Go to
Maintenance
and press
Maintenance
Position
, and needle will move to the right.
4) Carefully lift up the reaction chamber and also
rinsing chamber and dry up residual water in
this area.
5) Refit chambers, go to maintenance and press ”
Standby Position
” to let needle go
back to standby position.