background image

FLYABILITY

ELIOS 2 RAD

USER MANUAL

VERSION 1. 2

05.05.2022

FLYABILITY SA
+41 21 311 55 00 —

[email protected]

EPFL INNOVATION PARK BLDG C | 1015 LAUSANNE | SWITZERLAND

Содержание ELIOS 2 RAD

Страница 1: ...FLYABILITY ELIOS 2 RAD USER MANUAL VERSION 1 2 05 05 2022 FLYABILITY SA 41 21 311 55 00 INFO FLYABILITY COM EPFL INNOVATION PARK BLDG C 1015 LAUSANNE SWITZERLAND...

Страница 2: ...6 5 2 Out of range behavior 6 5 3 Settings Menu 7 6 Post flight analysis with Inspector 3 8 7 Calibration 10 7 1 Setting up RiumConfig 10 7 2 Calibrating the Elios 2 RAD sensor module 11 7 2 1 Connect...

Страница 3: ...of entering confined and cluttered environments Elios 2 RAD is equipped with an additional dosimetry sensor Figure 2 1 System overview The operator will enjoy full control of Elios through the Ground...

Страница 4: ...iation At its core the sensor module has an energy compensated Geiger M ller tube capable of detecting Cs137 and Co60 The sensor module is removable for calibration but can only be installed on the sp...

Страница 5: ...on 30s average 15 from 1 to 10mSv h 1sec sample rate 20 Sensitivity 3 5 uSv h limit of detection H 10 Saturation detection No 1 A calibration report is provided with the LOW dose detector 2 LOW dose D...

Страница 6: ...y is connected is displayed below the dose rate 2 SENSOR The type of sensor currently connected is indicated in the lower left corner of the plot area 3 This will indicate either Low medium or High 5...

Страница 7: ...ELIOS 2 RAD USER MANUAL Page 7 of 16...

Страница 8: ...gs Menu The cockpit settings menu contains the additional RAD SENSING menu tab This contains the following information and settings 1 Unit the unit in which the dose and dose rate are displayed this c...

Страница 9: ...sorbed until that point in the flight The user can choose whether the value is displayed in Sievert or Rem in the settings menu TRAJECTORY COLOR BAR 2 This menu allows you to change the color mapping...

Страница 10: ...ELIOS 2 RAD USER MANUAL Page 10 of 16...

Страница 11: ...ontact support flyability com and you will be given a new one free of charge 7 1 Setting up RiumConfig This software is developed by Icohup the manufacturer of the detector unit It can only be install...

Страница 12: ...tors which are named as follows in RiumConfig Low level X If the Elios 2 RAD sensor module is not connected to your PC at this point then you will get this message This message means No sensor verify...

Страница 13: ...type Icohup recommended calibration by using an irradiator allowing a homogeneous irradiation field at least 3 times larger than the sensor 7 2 3 Position The sensors must be positioned as follow the...

Страница 14: ...g range from 3 to 6 mSv h and 60 seconds irradiation time Once the calibration factor is adjusted we suggest validating an uncertainty below 10 from measurements at 20 50 and 80 of the global dose rat...

Страница 15: ...E Up to 9 min with RAD payload NON ROHS COMPLIANT Detector contains lead Applications exempted from the restriction in Article 4 1 specific to medical devices and monitoring and control instruments DI...

Страница 16: ...ex Switzerland Email support flyability com Tel 41 21 311 55 00 Content subject to change Find the latest version on www flyability com FLYABILITY SA 41 21 311 55 00 INFO FLYABILITY COM EPFL INNOVATIO...