S
ATURATION DETECTION AND BEHAVIOR
The saturation is not detected by the detector.
Above the maximum range the detector reaches its saturation point, above this value, the values will
continue to rise but they will no longer be accurate. When the ambient dose rate largely exceeds the
saturation level of the detector, the output measurement values will even decrease slightly. This fold-back
effect is not automatically detected.
5.3 Settings Menu
The cockpit settings menu contains the additional ‘RAD SENSING’ menu tab. This contains the
following information and settings:
1.
Unit, the unit in which the dose and dose rate are displayed, this can either be Sievert or Rem
2.
Measured Value, the dose rate currently measured by the detector, updated every 2 seconds
3.
Session Accumulated Dose, the total dose absorbed by the drone since the battery is connected
4. Lifetime Accumulated Dose, the total dose absorbed by the drone over its lifetime before this
session
5.
Connected sensor, this indicates the range of the detector currently connected.
ELIOS 2 RAD USER MANUAL
Page 8 of 16