Theory Of Operation

Introduction

3

 

 

 

 

3-1

Section 3 

Theory Of Operation

 

 
3.1 Introduction 

 

The user is cautioned about indiscriminately 
opening and disassembling the instrument.  The 
ionization chamber is pressurized and sealed at the 
factory, because the high impedance circuits of the 
ion chamber are easily contaminated with grease 
and dirt that produce electrical leakage.  The ion 
chamber cannot be disassembled for servicing   

 

The Model 451P is a pressurized ionization chamber instrument calibrated in exposure rate units of 
roentgens/hour (or sieverts/hour) for gamma and x-radiation in the energy range of 20 keV to 2 MeV.  The 
Model 451P responds to, but is not calibrated for beta radiation.  Beta energies that can be measured are 
above 1 MeV. 

The liquid crystal, supertwist display shows the radiation rate in digital and analog form with the range 
multiplier values also showing on the scale.  It is a lightweight electronic device that requires the 
computational capabilities of a microprocessor to make it operate.  It functions in a multiplex mode called 
quadriplex.  This mode uses four (4) back-planes to accommodate the 128 elements of the display. 

The microprocessor performs data collection, averaging, and multiplication by stored calibration factors, 
range changing, and battery check functions, in addition to driving the LCD.  Between computational 
periods, it “sleeps” in a low power mode to conserve battery power.  The microprocessor reads stored 
information from an electrically erasable memory, EEPROM, which is used by the program for calibration 
and display units.  The EEPROM will retain stored data when the instrument is OFF or when the batteries 
are removed.  Data can be entered into the EEPROM using the RS-232 port. 

Collection voltage for the ion chamber is approximately 105 V obtained from 5 lithium cells.  All internal 
power for the instrument is supplied by the 9-volt batteries. 

The digital and bargraph displays read directly.  The bargraph display update periods are listed in the 
table below.  The digital display updates at one second intervals nearest the current bar display update.  
The bargraph and digits display do not always show the same reading because the bargraph is faster 
than the digital update.  It is more convenient to watch the bargraph when the reading is changing quickly 
and to read the value of a slowly changing or static reading by looking at the digital display. 

The bargraph display is a digital presentation, programmed to appear as a linear analog meter display.  It 
is also referred to as the analog display throughout this manual. 

NOTE

Содержание Victoreen 451P

Страница 1: ...1P DE SI Ion Chamber Survey Meter Operators Manual February 2005 Manual No 1020101000 Rev 2 2004 2005 Fluke Corporation All rights reserved Printed in U S A All product names are trademarks of their respective companies ...

Страница 2: ...Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland Ohio 44139 440 498 2564 ...

Страница 3: ...1 2 1 External Controls 2 1 2 2 Installation 2 2 Section 3 Theory Of Operation 3 1 3 1 Introduction 3 1 3 2 Firmware 3 2 Section 4 Maintenance 4 1 4 1 Introduction 4 1 4 2 Routine Cleaning 4 1 4 3 Storage 4 1 4 4 Battery Replacement 4 2 4 5 High Voltage Battery Board Replacement 4 3 4 6 Replacement Parts Information 4 3 4 7 Recalibration and Service Information 4 4 ...

Страница 4: ......

Страница 5: ...measuring Units of measurement are displayed next to the 2 1 2 digit display LOW BAT and FREEZE will appear on the display when the instrument is operating in these modes External controls consist of an ON OFF button and a MODE button The unit is auto ranging and auto zeroing and has an auto ON backlight The backlight automatically comes on when the ambient light conditions fall below twilight lev...

Страница 6: ...tial is 105 Vdc and is derived from the 5 lithium cells The warm up time for an instrument that has been off 12 or more hours is about four minutes for readings less than 20 μR h in a 10 μR h or less background 1 3 Receiving Inspection Upon receipt of the unit 1 Check the shipping cartons s and their contents for in shipment damage If damage is evident file a claim with the carrier and contact Flu...

Страница 7: ... 1 2 digit display Analog Display the bargraph consists of 100 segments 2 inches long the scale has five major divisions the appropriate value for the operating range of the instrument will appear below the scale Digital Display the digital display is 2 1 2 digits followed by a significant zero digit depending on the operating range of the instrument The leading 1 2 digit is blank or a 1 or a 0 fo...

Страница 8: ...1 2 1 4 10 100 1000 10000 Energy keV Relative Response Side Face Front Model 451P DE SI Typical Energy Dependence Dose equivalent units were determined using conversion coefficients at H 10 found in ICRU 47 April 1992 0 0 2 0 4 0 6 0 8 1 1 2 1 4 10 100 1000 10000 Energy keV Relative Response Side Front ...

Страница 9: ...General Information Specifications 1 1 5 Survey Meter Angular Response ...

Страница 10: ...Victoreen 451P 451P DE SI Operators Manual ...

Страница 11: ...decreases through the lower ranges to a reading of less than 50 μR h within 120 seconds When the element s in the display turn off with the exception of those necessary for normal operation the user can begin the measurement process A one minute warm up period is recommended before making a measurement If the unit experiences an extreme temperature change e g moving inside from outside a longer wa...

Страница 12: ... configured as a toggle to display exposure rate integrated exposure If the MODE button is pressed within the 30 second initialization period of the unit the display will read 0 When integration starts 0 0 μR will be displayed Toggle MODE to read exposure rate as required To reset the integration display toggle the display from the Rate mode to the Integrate mode Keep the MODE button pressed for a...

Страница 13: ...ge CO to be displayed on the 451P 1 Be sure the computer communication port is active 2 The 451 communication transmitter RCR pair is active when it detects a negative mark voltage at the receiver input CO shall appear on the 451 display 3 Press space bar twice If there is still a problem consult the Fluke Biomedical Radiation Management Services Repair Coordinator for further instructions ...

Страница 14: ...Victoreen 451P 451P DE SI Operators Manual ...

Страница 15: ...llection averaging and multiplication by stored calibration factors range changing and battery check functions in addition to driving the LCD Between computational periods it sleeps in a low power mode to conserve battery power The microprocessor reads stored information from an electrically erasable memory EEPROM which is used by the program for calibration and display units The EEPROM will retai...

Страница 16: ...t that the calibration coefficients track from range to range because an oscillatory condition can occur if the calibration on a given range is low and the coefficient for the next more sensitive range is high 3 2 Firmware The program in the 451 ROM is proprietary to Fluke Biomedical Radiation Management Services The firmware version appears in the digital part of the LCD display prior to the all ...

Страница 17: ...outlined in ANSI N45 2 2 1972 Section 6 1 2 2 The storage area shall comply with ANSI N45 2 2 1972 Section 6 2 Storage Area paragraphs 6 2 1 through 6 2 5 Housekeeping shall conform to ANSI N45 2 3 1972 Level B components must be stored within a fire resistant tear resistant weather tight enclosure in a well ventilated building or equivalent Storage of Fluke Biomedical Radiation Management Service...

Страница 18: ...he LO BAT message appears The instrument will function on one battery for approximately 100 hours allowing replacement of one battery at a time if the instrument must remain operational during battery changeover Regular or alkaline batteries can be used for replacement purposes Both 9 volt batteries are located in the rear of the unit and are easily accessible and replaceable with the battery cove...

Страница 19: ... 6 Secure the battery board to the survey meter assembly with the two screws lock washers and aluminum spacers removed in step 4 7 Secure the case top to the case bottom with the gasket in between using the four screws removed in step 2 8 Replace the 9 volt batteries being sure to observe proper polarity 4 6 Replaceable Parts Information Fluke Biomedical Radiation Management Services maintains a c...

Страница 20: ...ration or repair we request that you consult the Fluke Biomedical Radiation Management Services Repair Coordinator at 800 850 4606 440 498 2564 or Fax 440 542 3682 More information concerning the operation application or service of your instrument may be obtained from the applications engineer at the numbers listed above ...

Страница 21: ......

Страница 22: ...Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland Ohio 44139 440 498 2564 ...

Отзывы: