Thermalert
®
4.0 Series
Users Manual, Rev. 2.4, Jul 2021
106
14 Maintenance
Our sales representatives and customer service staff are always at your disposal for questions regarding
applications, calibration, repair, and solutions to specific problems. Please contact your local sales representative
if you need assistance. In many cases, problems can be solved over the telephone. If you need to return equipment
for servicing, calibration or repair, please contact our Service Department before shipping. Phone numbers are
listed at the beginning of this document.
14.1 Troubleshooting Minor Problems
Table 14-1: Troubleshooting
Symptom
Probable Cause
Solution
No output
No power to instrument
Check the power supply
Erroneous temperature
Faulty sensor cable
Verify cable continuity
Erroneous temperature
Field of view obstruction
Remove the obstruction
Erroneous temperature
Window lens
Clean the lens
Erroneous temperature
Wrong emissivity
Correct the setting
Temperature fluctuates
Wrong signal processing
Correct Peak/Valley Hold or Average settings
Temperature fluctuates
No ground for the instrument Check wiring / grounding
14.2 Fail-Safe Operation
The Fail-Safe system is designed to alert the operator and provide a safe output in case of any system failure.
Basically, it is designed to shut down the process in the event of a set-up error, system error, or a failure in the
sensor electronics.
Warning
The Fail-Safe circuit should never be relied on exclusively to protect critical heating processes.
Other safety devices should also be used to supplement this function!
When an error or failure does occur, the output circuits automatically adjust to the lowest or highest preset level.
The following table shows the corresponding analog values and the error code transmitted over the digital
communication.
Table 14-2: Error Codes for Outputs
Symptom
0 to 10 V 0 to 20 mA
4 to 20 mA
Relay
Measured temperature over range*
10 V
21 mA
21 mA
alarm state
Measured temperature under range*
0 V
0 mA
approx. 3.5 mA alarm state
* related to zoomed temperature range
Table 14-3: Error Codes via Digital Communication
Output
Error Code Description
EHHH
Measured temperature over range
EUUU
Measured temperature under range
EIHH
Internal temperature over range
EIUU
Internal temperature under range
www.
.com
1.800.561.8187
Содержание Thermalert 4.0 Series
Страница 131: ...Thermalert 4 0 Series Users Manual Rev 2 4 Jul 2021 132 www com information itm com 1 800 561 8187...
Страница 133: ...Thermalert 4 0 Series Users Manual Rev 2 4 Jul 2021 134 www com information itm com 1 800 561 8187...
Страница 134: ...Appendix IECEx Certificate of Conformity 135 15 www com information itm com 1 800 561 8187...
Страница 135: ...Thermalert 4 0 Series Users Manual Rev 2 4 Jul 2021 136 www com information itm com 1 800 561 8187...