Battery Type and Life:  AA (2); 400 hours continuous with
alkaline; 200 hours continuous with zinc chloride

Shock, Vibration:  1 m drop at 15 

°

C to 35 

°

C (59 

°

F to 95 

°

F)

per ANSI/ISA-S82.01-1994 and EN 61010-1: 1993. Sinusoidal
vibration per MIL-PRF-28800F for a Class 2 instrument
(5 Hz to 55 Hz, 3 g maximum).

Surge Protection:  T5-600 6 kV per IEC 1010-1, 1990-09;

T5-1000 8 kV per IEC 1010-1, 1990-09

Enclosure Rating:  IP 52 per IEC 529, no vacuum applied

RF Field Specification:  0.5 % full scale + (specified accuracy)
at 3 V / m; Amps: >100 MHz, performance unspecified.

Safety:  Complies with ANSI/ISA-S82.01-94 for use in
Overvoltage category III (CAT III) locations, UL3111, CSA/CAN
C22.2 No.1010.1-92, and EN61010-1: 1993.

EMC:  EN 50081-1, EN 50082-1

Certifications:

Accuracy is specified for 1 year after calibration, at 18 

°

C to 28

°

C (64 

°

F to 82 

°

F) with relative humidity to 90 %. AC

conversions are ac-coupled, average responding, and calibrated
to the rms value of a sine wave input.
Accuracy specifications are given as follows:

±

 ( [ % of reading] + [number of least significant digits] )

Temperature coefficient of 0.1 x (specified accuracy) / 

°

C for

<18 

°

C or >28 

°

C (< 64.4 

°

F or > 82.4 

°

F)

Function

T5-600

Range

T5-1000

Range

Resolution

Accuracy

K

600 V rms

1000 V rms

1 V

±

 (1.5 % + 2 digits)

L

600 V

1000 V

1 V

±

 (1 % + 1 digit)

?

100.0 A

100.0 A

0.1 A

±

 (3 % + 3 digits)

e

1000 

1000 

±

 (1 % + 2 digits)

Input Protection

Function

Input Impedance

(nominal)

T5-600

T5-1000

K

1 M

, <100 pF ac-coupled

600 V rms

1000 V rms

L

1 M

, <100 pF

600 V rms

1000 V rms

e

>2.6 k

600 V rms

1000 V rms

Function

Open Circuit Test Voltage

Short Circuit Current

e

2.4 V dc (nominal)

<600 

µ

A