Measuring AC Current  (A)

0.5"(12.9 mm)

maximum

A

AC

ex1.eps

  Range:  0 A to 100.0 A;

45 Hz to 66 Hz

  Remove probes from test

points.

 Place conductor anywhere

within the sensor zone
(shaded area shown below).

OK

OK

OK

OK

Sensor 
Zone

ex5.eps

Display Hold 

To hold the reading on the display, press and release HOLD.
The voltage indicator LED (

) continues to operate.

To exit the display hold mode, press and release HOLD again, or
turn the rotary switch to a new position.

Automatic Power Down

The tester turns off automatically if you do not turn the rotary
switch or press HOLD for approximately 34 minutes. To resume
operation, turn the tester off, wait two seconds,  then turn the
tester on.

Low Battery Indicator  

W

 Warning

To avoid false readings, which could lead to
possible electric shock or personal injury, replace
the battery as soon as the low battery indicator
(

M

) appears.