Fluke 192/196/199
User Manual
2-4
To move the trace:
1
Move the trace up or down until
the ground marker hits the top -
or bottom grid line.
When moving down, the following banner will show up:
Figure 2-3. Extended offset banner
When moving up a similar banner shows up.
Oscilloscopes and dynamic range
To prevent waveforms from being distorted, in particular
signals with large very steep edges, the signal should stay
within the dynamic range, just as with any Oscilloscope.
The dynamic range limits are the same as the trace offset
limits given in the section Specifications below.
In practice you must take care when using high time base
speeds when looking at signals that have large very steep
edges. They can be distorted when they jump from
outside the dynamic range directly to the visible area of
the screen.
Figure 2-4 shows a pulse wave within the dynamic range
(left), and a pulse wave that is partially outside the
dynamic range (right). The example, particularly the
16 div trace offset, applies to one of the 4 most sensitive
ranges.