Using the Medical Functions
mAs Measurements
2
2-3
From step 2, you can also proceed as follows to do mAs
measurements on input A:
3
In Scope mode display the
INPUT
A
key labels.
4
Open the
Probe on A
menu.
5
Select
Probe Type: Current
,
then
accept.
6
Select for example
1 V/A
, then
accept.
7
Display the cursor key labels (B-
and C-versions are different).
8
Press to highlight
. Observe
that two
vertical
cursors are
displayed.
9
Press to highlight mAs.
10
Highlight the left or right cursor.
11
Move the highlighted cursor to the
desired position on the waveform.
mVs measurements
When you select
Voltage
in the
Probe on A (B)
menu,
the F3 key label shows
mVs
.
mWs measurements
When you are in the Math AxB mode, and you select in
the
Probe on A (B)
menu for one channel
Voltage
and for
the other channel
Current
, the F3 key label will show
mWs.