Fluke 190B
/
C Medical
User Manual
2-2
3
Display the
S
COPE
key labels
4
Open the Reading 1 menu
5
Select
on A
, then accept
6
Select
mAs,
then accept. This
opens the
Current Probe
menu.
7
Select for example
1 V/A
, then
accept.
After closing the information banner the screen shows
(see figure 2-2) :
•
Top left reading,
mA
(or A,
µ
A, etc.):
the difference between the current amplitude measured
on the left cursor and measured on the right cursor:
•
Top right reading,
mAs
(or As,
µ
As, etc.):
the integrated current over the time between the
cursors.
Figure 2-2. mAs measurement screen
Note: the B versions F3 key label does not show
RMS
as
these versions do not provide RMS measurements
on the trace section between the cursors.
8
Highlight the left or right cursor.
9
Move the highlighted cursor to the
desired position on the waveform.