ScopeMeter® Test Tool
Performance Verification
31
Continuity Function Test
To test the continuity function:
1. Press
M
.
2. Press
1
to open the Measurement menu and select
Continuity
.
3. Connect the Test Tool to the 5502A. See
.
4. Set the 5502A to 20
Ω
. Use the 5502A “COMP OFF” mode.
5. Listen to hear that the beeper is on.
6. Set the 5502A to
80
Ω
.
7. Listen to hear that the beeper is off.
When you are finished, set the 5502A to Standby.
Diode Test Function Test
To test the diode function:
1. Select the Test Tool setup.
2. Press
M
.
3. Press
1
to open the Measurement menu and select
Diode
.
4. Connect the Test Tool to the 5502A. See
.
5. Set the 5502A to
1 k
Ω
. Use the 5502A “COMP OFF” mode.
6. Observe the main reading and check that it is within
0.4 V
and
0.6 V
.
7. Set the 5502A to
1 V dc
.
8. Observe the main reading and check that it is within
0.975 V
and
1.025 V
.
When you are finished, set the 5502A to Standby.
Ohms Measurements Test
To test the Ohms measurement accuracy:
1. Connect the Test Tool to the 5502A. See
.