190 Series III
Service Manual
28
External Trigger Level Test
Note
The external trigger level test is for the models 190-062, 190-102, 190-202, 190M-2,
and 190-502.
To test the external trigger level:
1. Connect the Test Tool to the 5502A. See
.
Figure 10. Test Tool Meter/Ext Input to 5502A Normal Output
2. Reset the Test Tool to select the Test Tool setup.
3. Press
T
.
4. Use
4
to select the
TRIGGER OPTIONS...
menu.
5. Select
On Edges...
from the TRIGGER OPTIONS menu.
6. Press
E
.
7. Select
Update: Single Shot
E
,
Trigger Filter: Noise Reject
E
,
NCycle: Off
E
.
8. Use
1
(EDGE TRIG) to select Ext.
9. Use
2
(SLOPE) to select positive slope triggering (
).
10. Use 3 (Ext LEVEL) to select
1.2 V
.
11. Set the 5502A to source
0.4 V dc
.
12. Verify that no trace is shown on the Test Tool display and that the status line at the display
top shows
SINGLE MANUAL
or
SINGLE WAITING
.
If the display shows the trace and status as
SINGLE HOLD
, press
H
to re-arm the Test
Tool for a trigger.
13. Set the 5502A to source
1.7 V
.
14. To verify that the Test Tool is triggered, check that the trace becomes visible. To repeat
the test, start at step 3.
Set the 5502A to Standby.