190 Series III
Service Manual
32
Figure 11. Test Meter Tool Input to 5502A Normal Output 4-Wire
2. Select the Test Tool setup.
3. Press
M
.
4. Press
1
to open the Measurement menu and select
Ohms
.
5. Press
J
to select AUTO ranging.
6. Set the 5502A to source the appropriate resistance value for the first test point in
7. Use the 5502A “COMP 2 wire” mode for the verifications up to and including 50 k
Ω
. For the
higher values, the 5502A will turn off the “COMP 2 wire” mode.
8. Observe the reading and check that it is within the range shown under the appropriate
column.
9. Continue through the test points.
When you are finished, set the 5502A to Standby.
Table 13. Resistance Measurement Verification Points
5502A Output
Meter Reading (Comp 2-Wire)
0
Ω
0.0 to 0.5 (COMP 2 wire)
400
Ω
397.1 to 402.9 (COMP 2 wire)
4 k
Ω
3.971 to 4.029 (COMP 2 wire)
40 k
Ω
39.71 to 40.29 (COMP 2 wire)
400 k
Ω
397.1 to 402.9 (off)
4 M
Ω
3.971 to 4.029 (off)
30 M
Ω
29.77 to 30.23 (off)
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
USE STACKABLE TEST LEADS
AS SUPPLIED WITH CALIBRATOR
A
B
METER
2 CHANN. + METER