Calibration Procedure
–
Remove the cover from the counter.
–
Allow the MTCXO to adapt the new ambient tempera-
ture. (See ‘Preparations’.)
–
Connect the 10 MHz reference to Input-A.
–
Switch ON the counter.
–
Adjust the sensitivity control so that the counter counts
properly.
–
Hold down the CALIB-button, on the main printed-circuit
board in the counter, and press the Reset-button.
–
Wait about 20 seconds, until the display shows
10.0000000 MHz. Now the oscillator is calibrated.
–
Switch OFF the counter and disconnect the 10 MHz ref-
erence.
–
Fit the cover.
CALIB-button
Figure 6
Location of the CALIB-Button.