Trigger Setup
Use Trigger to determine when the Analyzer should start displaying waveforms on the screen so that
the waveforms displayed by each refresh can be aligned. Once the set trigger condition is met, the
screen is triggered, the waveform displayed on the screen is aligned at that point.
The Analyzer offers a variety of trigger methods to support multiple trigger sources and trigger modes.
Enter the Trigger Setup Menu:
•
From the main screen of Scope select
C
(
Trigger
).
The options are as shown in
Table 21. Trigger Setup
Source
A
Source
Push
A
(Source) to cycle through the current trigger sources.
The currently selected trigger source is shown under the “
Source
” label with a
color block of the same color as the waveform.
The selectable trigger sources are the selected signal in the “
Signal
” menu,
which is the signal displayed on the “
Vertical Axis Scale
” of the screen, see
Select the Waveform to Display
Trigger Option
B
Trigger Option
Configure trigger mode and trigger edge.
-
Method
: optioins are “
Auto
”, “
Normal
” and “
Single Shot
”. The default is
“
Auto
” mode.
-
Slope
: options are Positive,
Negative and Dual. The default
is Positeve.
�
�
D
E
Trigger Level
The up and down arrow keys increase or decrease the trigger level. Each time
the key is pushed, the trigger mark on the screen moves in the up/down
direction.
Note
If the trigger level is not set properly, under the “Auto” mode, the starting point of the
waveform displayed (signal level at the left side of the screen) may be unstable, it will
jitter left and right; the waveform may not be displayed at all under “Normal” mode.