DO NOT
use this unit in environments other than those listed in the
user’s manual.
•
DO NOT
use this unit for any application other than calibration work.
•
Follow all safety guidelines listed in the user’s manual.
•
Calibration Equipment should only be used by Trained Personnel.
1.2.2
CAUTIONS
To avoid possible damage to the instrument, follow these guidelines.
•
Operate the instrument in room temperatures between 5-50°C (41-122°F).
Allow sufficient air circulation by leaving at least 6 inches of space be-
tween the instrument and nearby objects. Overhead clearance needs to al-
low for safe and easy insertion and removal of probes for calibration.
•
The dry-well is a precision instrument. Although it has been designed for
optimum durability and trouble free operation, it must be handled with
care. The instrument should not be operated in excessively wet, oily,
dusty, or dirty environments. It is important to keep the wells of the in-
strument clean and clear of any foreign matter. Do not operate near flam-
mable materials.
•
DO NOT
use fluids to clean out the well.
•
If a mains supply power fluctuation occurs, immediately turn off the in-
strument. Power bumps from brown-outs and black-outs could damage
the instrument. Wait until the power has stabilized before re-energizing
the instrument.
1.3
Authorized Service Centers
Please contact one of the following authorized Service Centers to coordinate
service on your Hart product:
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: [email protected]
Fluke Nederland B.V.
Customer Support Services
Science Park Eindhoven 5108
3
1 Before You Start
Authorized Service Centers