3
Specifications and Warranty
3.1
Specifications
See Table 2.
Temperature Range
–200°C to 156°C
Entire Assembly Range
–200°C to 156°C
Accuracy (Includes calibration
uncertainty and short-term
stability)
±0.05°C (full range)
Long-term Drift (°C)
<±0.01 at 0°C per year, when used periodically at 200
Hysteresis (°C)
<0.01
Temperature Coefficient
0.003925 ohms/ohms/°C
Size (O.D.)
0.25" (6.35 mm)
Sheath Length
6" (152.4 mm)
Sheath Material
Inconel 600
Cable length
20 ft. (609.6 cm)
Lead Wires
22 AWG/ Teflon insulated cable, 4-wire
3.2
Warranty
Hart Scientific, Inc. (Hart) warrants this product to be free from defects in ma-
terial and workmanship under normal use and service for a period as stated in
our current product catalog from the date of shipment. This warranty extends
only to the original purchaser and shall not apply to any product which, in
Hart’s sole opinion, has been subject to misuse, alteration, abuse or abnormal
conditions of operation or handling.
Software is warranted to operate in accordance with its programmed instruc-
tions on appropriate Hart products. It is not warranted to be error free.
Hart’s obligation under this warranty is limited to repair or replacement of a
product which is returned to Hart within the warranty period and is determined,
upon examination by Hart, to be defective. If Hart determines that the defect or
malfunction has been caused by misuse, alteration, abuse or abnormal condi-
tions or operation or handling, Hart will repair the product and bill the pur-
chaser for the reasonable cost of repair.
5623A
7
3 Specifications and Warranty
Table 2
Specifications