Description
Command
Example
Response
Format
Response
Example
Range
Read probe wires
w[ires]
wi
wi: 9
wi: 4
Set probe wires
w[ires]=2/3/4
wi=4
2,3,4
Read probe calibration date
pc[al]
pcal
pc: yyyy-mm-dd
pc: 1999-03-28
Set probe calibration date
pc[al]=yyyy-mm-dd
pc=1999-03-28
1971-01-01 to
2099-12-31
Read probe due date
pd[ue]
pdue
pd: yyyy-mm-dd
pd: 1999-03-28
Set probe due date
pd[ue]=yyyy-mm-dd
pd=1999-03-28
1971-01-01 to
2099-12-31
Read filter
fi[lter]
fi
fi: 99.999999
fi: 0.5
Set filter
fi[lter]=n
fi=0
0.0 to 60.0
Read low range calibration
*c1
*c1
*c1: 9.999
*c1: -0.0121
Set low range calibration
*c1=n
*c1=-0.012
–1.999 to
1.999
Read high range calibration
*c2
*c2
*c2: 9999.9
*c2: -0.9
Set high range calibration
*c2=n
*c2=-0.9
–1980.0 to
1980.0
Read meter calibration date
mc[al]
mcal
mc: yyyy-mm-dd
mc: 1999-03-28
Set meter calibration date
mc[al]=yyyy-mm-dd
mc=1999-03-28
1971-01-01 to
2099-12-31
Read meter due date
md[ue]
mdue
md: yyyy-mm-dd
md: 1999-03-28
Set meter due date
md[ue]=yyyy-mm-dd
md=1999-03-28
1971-01-01 to
2099-12-31
Legend:
[] Optional Command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
51
8 Communications Interface
Communication Command List