DPM2 Plus
Operators Manual
2
Safety Information
W
Caution
To avoid personal injury and possible damage to the
Meter:
•
Use the Meter only with gases and liquids found in Table
2
of this manual.
•
Use appropriate safety precautions when working with
hazardous liquids and gases.
•
Be aware of the direction of gas flow to avoid hazardous
conditions during bleed-off procedure.
•
For pressures above 5 psi and vacuum pressures of -5 psi,
use only securely clamped reinforced tubing.
General Specifications
Power Requirements
............................................ One 9-volt battery, NEDA 6LR61,
IEC 1604A
Battery Life
............................................................ 60 hours of continuous use
Operating Temperature
........................................ 0
°
C to 30
°
C (32
°
F to 86
°
F)
Overpressure
........................................................ Not to exceed 110 psi
Pressure Connection
........................................... Male Luer–Lok
Ranges
................................................................... -698 to +802 mmHg
-949 to +1090 cm H
2
O
-374 to +429 in H
2
O
-13.50 to +15.50 psi
-13.5 to +100.0 psi
Accuracy
............................................................... ±1 % of full scale at all ranges
Voltage Output
...................................................... 0.1 V/psi on all ranges except 100 psi,
which is 0.01 V/psi
Voltage Output Accuracy
..................................... 5 % of range
Display
................................................................... 0.5–in LCD with
LO BATT
Negative polarity indication
Dimensions
........................................................... 14.61
×
9.14
×
4.83 cm
(5.75
×
3.6
×
1.9 in)