Calibration Manual 

 ii 

 

Содержание CNX i3000

Страница 1: ...pecifications are subject to change without notice All product names are trademarks of their respective companies CNX i3000 Wireless iFlex Calibration Manual rie enia na presn meranie Elso Philips Ser...

Страница 2: ...r nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB...

Страница 3: ...pment 5 Performance Tests 5 Test the Display 5 Backlight 6 Keypad Test 6 AC Current Test 6 Before Calibration Adjustment 9 Maintenance Mode 9 Password Entry 9 Change the Password 9 Restore the Default...

Страница 4: ...CNX i3000 Calibration Manual ii...

Страница 5: ...iii List of Tables Table Title Page 1 Symbols 3 2 Required Equipment 5 3 Performance Tests 6 4 Calibration Adjustment 11 5 User Replaceable Parts 13...

Страница 6: ...CNX i3000 Calibration Manual iv...

Страница 7: ...s Figure Title Page 1 All Segments of the Display 5 2 Performance Test Connections for Applied Voltages 7 3 Performance Test Connections for Applied Current 8 4 Calibration Password Reset 10 5 Battery...

Страница 8: ...CNX i3000 Calibration Manual vi...

Страница 9: ...lowing telephone numbers Technical Support USA 1 800 44 FLUKE 1 800 443 5853 Calibration Repair USA 1 888 99 FLUKE 1 888 993 5853 Canada 1 800 36 FLUKE 1 800 363 5853 Europe 31 402 675 200 Japan 81 3...

Страница 10: ...you operate the Product Replace the batteries when the low battery indicator shows to prevent incorrect measurements Have an approved technician repair the Product Use only specified replacement parts...

Страница 11: ...he low voltage MAINS installation CAT III Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the building s low voltage MAINS installation CAT...

Страница 12: ...000 m 6561 68 ft Storage Altitude 12 000 m 39370 079 ft Temperature Coefficient Add 0 1 X specified accuracy C 18 C or 28 C Add 0 1 X specified accuracy F 64 4 F or 82 4 F Safety Rating IEC 61010 1 60...

Страница 13: ...trical shock fire or personal injury do not perform the performance test procedures unless the Product is fully assembled The performance tests verify the full operation of the Product and measure the...

Страница 14: ...battery is good and replace it if necessary See Battery Replacement 3 Warm up the Calibrator as necessary Refer to its specifications 4 Let the temperature of the unit under test UUT become stable to...

Страница 15: ...2 10 8 0 3 A 500 Hz 0 1 14 05 15 95 1 180 mV 500 Hz 0 1 581 5 618 5 1 300 mV 500 Hz 1 965 1035 1 750 mV 500 Hz 1 2420 2580 Freq R coil 5 Hz 0 3 A 0 1 0 5 4 5 5 5 50 Hz 0 4 A 0 1 49 3 50 8 500 Hz 0 325...

Страница 16: ...V RMS MAX 20V PK MAX 1V PK MAX 20V PK MAX NORMAL AUX TRIG OUT SCOPE V RTD A SENSE AUX V 150V PK MAX 20V PK MAX 50 Turn Current Coil CNX i3000 Conductor hca009 eps Figure 3 Performance Test Connections...

Страница 17: ...g the procedures packaging and bench techniques that are recommended The following practices should be followed to minimize damage to S S static sensitive devices 1 MINIMIZE HANDLING 2 KEEP PARTS IN O...

Страница 18: ...H OPEN EDGE CONNECTOR EXCEPT AT STATIC FREE WORK STATION PLACING SHORTING STRIPS ON EDGE CONNECTOR HELPS PROTECT INSTALLED S S DEVICES 9 HANDLE S S DEVICES ONLY AT A STATIC FREE WORK STATION 10 ONLY A...

Страница 19: ...choice The subsequent flashes 5 Do steps 3 and 4 again to enter the subsequent digits of the four number password 6 When all of the correct digits are entered push to confirm the input If the correct...

Страница 20: ...the bottom case screws Two of the screws are inside of the battery door 3 Keep the pca in the top case 4 Apply 3 0 V across the battery contacts on the pca Note the polarity that is shown in Figure 4...

Страница 21: ...calibration step number advances before you change the calibrator source Some adjustment steps can take several seconds to execute before the Product goes to the subsequent step Set the Calibrator to...

Страница 22: ...er test Remove batteries to prevent battery leakage and damage to the Product if it is not used for an extended period Be sure that the battery polarity is correct to prevent battery leakage To change...

Страница 23: ...e shown in Table 5 Table 5 User Replaceable Parts Fluke Part Number Description Qty 4130305 FLK 3000 2003 DOOR BATTERY 1 3676410 FLUKE I2500 10 IFLEX 2500A PROBE 10IN 1 1881997 TPAK meter hanging kit...

Страница 24: ...CNX i3000 Calibration Manual 14 rie enia na presn meranie http www elso sk...