![Fluke CNX a3000 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/fluke/cnx-a3000/cnx-a3000_calibration-manual_2309845020.webp)
CNX a3000
Calibration Manual
10
Restore the Default Password
If the calibration password is lost, the default password (1234) can be manually restored
with the subsequent steps:
Warning
To prevent electric shock or personal injury, remove all input
signals before you open the Product.
1.
Remove the Product battery door. See “Battery Replacement”.
2.
With a Phillips screwdriver, remove the bottom case screws. Two of the screws are
inside of the battery door.
3.
Keep the pca in the top case.
4.
Apply 3.0 V across the battery contacts on the pca. Note the polarity that is shown in
Figure 4.
5.
Push
on the front of the Product.
6.
Short across the CAL keypad on the pca. See Figure 4. The default password is now
restored.
7.
Remove the 3.0 V supply and replace the bottom case, batteries, and battery door.
Short across calibration
key pads to reset default
password
Battery
contacts
hby08.eps
Figure 4. Calibration Password Reset
riešenia na presné meranie
http://www.elso.sk
Содержание CNX a3000
Страница 4: ...CNX a3000 Calibration Manual ii...
Страница 6: ...CNX a3000 Calibration Manual iv...
Страница 8: ...CNX a3000 Calibration Manual vi...
Страница 24: ...CNX a3000 Calibration Manual 14 rie enia na presn meranie http www elso sk...