Fluke CNX a3000 Скачать руководство пользователя страница 1

 

November 2012 

© 2012 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. 
All product names are trademarks of their respective companies. 

CNX a3000 

Wireless AC Clamp 

 
 
 

Calibration Manual 

 

riešenia na presné meranie™

Elso Philips Service; tel: +421 32 6582410
email: 

[email protected]

; web: 

www.elso.sk

Содержание CNX a3000

Страница 1: ...cifications are subject to change without notice All product names are trademarks of their respective companies CNX a3000 Wireless AC Clamp Calibration Manual rie enia na presn meranie Elso Philips Se...

Страница 2: ...r nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB...

Страница 3: ...pment 6 Performance Tests 6 Test the Display 6 Backlight 7 Keypad Test 7 AC Current Test 7 Before Calibration Adjustment 9 Maintenance Mode 9 Password Entry 9 Change the Password 9 Restore the Default...

Страница 4: ...CNX a3000 Calibration Manual ii...

Страница 5: ...iii List of Tables Table Title Page 1 Symbols 4 2 Required Equipment 6 3 Performance Tests 7 4 Calibration Adjustment 11 5 User Replaceable Parts 13...

Страница 6: ...CNX a3000 Calibration Manual iv...

Страница 7: ...v List of Figures Figure Title Page 1 Tactile Barrier 3 2 All Segments of the Display 6 3 Performance Test Connections 8 4 Calibration Password Reset 10 5 Battery Replacement 13...

Страница 8: ...CNX a3000 Calibration Manual vi...

Страница 9: ...llowing telephone numbers Technical Support USA 1 800 44 FLUKE 1 800 443 5853 Calibration Repair USA 1 888 99 FLUKE 1 888 993 5853 Canada 1 800 36 FLUKE 1 800 363 5853 Europe 31 402 675 200 Japan 81 3...

Страница 10: ...if it operates incorrectly The battery door must be closed and locked before you operate the Product Replace the batteries when the low battery indicator shows to prevent incorrect measurements Have a...

Страница 11: ...ore use if the battery leaks Be sure that the battery polarity is correct to prevent battery leakage Batteries contain hazardous chemicals that can cause burns or explode If exposure to chemicals occu...

Страница 12: ...ement Category IV is applicable to test and measuring circuits connected at the source of the building s low voltage MAINS installation Conforms to European Union directives Conforms to relevant North...

Страница 13: ...rating Altitude 2 000 m 6561 68 ft Storage Altitude 12 000 m 39370 079 ft Ingress Protection IP rating IP30 Radio Frequency Certification FCC Part 15 Subpart C Sections 15 207 15 209 15 249 FCCID T68...

Страница 14: ...nce test procedures unless the Product is fully assembled The performance tests verify the full operation of the Product and measure the accuracy of each function against Product specifications If the...

Страница 15: ...t the temperature of the UUT become stable to room temperature To do the ac current test 1 Connect the Calibrator A ac output and ground to the 50 Turn Coil See Figure 3 2 Apply the input level for ea...

Страница 16: ...RMS MAX 20V RMS MAX 20V PK MAX 1V PK MAX 20V PK MAX NORMAL AUX TRIG OUT SCOPE V RTD A SENSE AUX V 150V PK MAX 20V PK MAX 5500A 50 Turn Current Coil Conductor hby07 eps Figure 3 Performance Test Connec...

Страница 17: ...g the procedures packaging and bench techniques that are recommended The following practices should be followed to minimize damage to S S static sensitive devices 1 MINIMIZE HANDLING 2 KEEP PARTS IN O...

Страница 18: ...H OPEN EDGE CONNECTOR EXCEPT AT STATIC FREE WORK STATION PLACING SHORTING STRIPS ON EDGE CONNECTOR HELPS PROTECT INSTALLED S S DEVICES 9 HANDLE S S DEVICES ONLY AT A STATIC FREE WORK STATION 10 ONLY A...

Страница 19: ...choice The subsequent flashes 5 Do steps 3 and 4 again to enter the subsequent digits of the four number password 6 When all of the correct digits are entered push to confirm the input If the correct...

Страница 20: ...the bottom case screws Two of the screws are inside of the battery door 3 Keep the pca in the top case 4 Apply 3 0 V across the battery contacts on the pca Note the polarity that is shown in Figure 4...

Страница 21: ...quent calibration step Note After you push wait until the calibration step number advances before you change the calibrator source Some adjustment steps can take several seconds to execute before the...

Страница 22: ...r test Remove batteries to prevent battery leakage and damage to the Product if it is not used for an extended period Be sure that the battery polarity is correct to prevent battery leakage To change...

Страница 23: ...placeable Parts User replaceable parts are shown in Table 5 Table 5 User Replaceable Parts Fluke Part Number Description Qty 4108300 FLK A3000 2003 DOOR BATTERY 1 376756 Battery AA 1 5 V NEDA 15 A IEC...

Страница 24: ...CNX a3000 Calibration Manual 14 rie enia na presn meranie http www elso sk...