9116A Furnace
User’s Guide
44
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Set program control mode:
pc=g[o]/s[top]/c[ont]
GO or STOP or
CONT
Start program
pc=g[o]
pc=g
Stop program
pc=s[top]
pc=s
Continue program
pc=c[ont]
pc=c
Read program function
pf
pf
pf: 9
pf: 3
Set program function to
n
pf=
n
pf=2
1 to 4
Configuration Menu
Probe Menu
Read R0 calibration parameter
r[0]
r
r0: 999.999
r0: 100.578
Set R0 calibration parameter to
n
r[0]=n
r=100.324
98.0 to 104.9
Read ALPHA calibration
parameter
al[pha]
al
al: 9.9999999
al: 0.0038573
Set ALPHA calibration parameter
to
n
al[pha]=n
al=0.0038433
.00370 to
.00399
Read DELTA calibration
parameter
de[lta]
de
de: 9.99999
de: 1.46126
Set DELTA calibration parameter
to
n
de[lta]=
n
de=1.45
0.0 to 2.9
Operating Parameters Menu
Set temperature units:
u[nits]=c/f
C or F
Set temperature units to Celsius
u[nits]=c
u=c
Set temperature units to
Fahrenheit
u[nits]=f
u=f
Read cutout mode
cm[ode]
cm
cm: {xxxx}
cm: AUTO
Set cutout mode:
cm[ode]=r[eset]/a[uto]
RESET or
AUTO
Set cutout to be reset manually
cm[ode]=r[eset]
cm=r
Set cutout to be reset
automatically
cm[ode]=a[uto]
cm=a
Read approach setting
ap[proach]
ap
ap:9
ap:5
Set approach setting to
n
degrees
ap[proach]=
n
ap=15
0 to 20°C
Read stability
ts
ts
ts:9.9
ts:0.5
Set soak stability to
n
degrees
ts=
n
ts=.1
.01 to 4.99°C
Serial Interface Menu
Read serial sample setting
sa[mple]
sa
sa: 9
sa: 1
Digital Interface Command Summary continued