87V MAX
Calibration Information
12
Testing Meter Accuracy
To test the accuracy of the Meter, perform the steps in Table 2.
Table 2. Accuracy Tests
Step
Test
Function
Range
5520A Output
Display Reading
1
AC Volts
600 mV
60 mV, 60 Hz
59.2 to 60.8
2
600 mV
330 mV, 60 Hz
327.3 to 332.7
3
600 mV
600 mV, 13
kHz
586.0 to 614.0
4
6 V
675 mV, 60 Hz
0.666 to 0.684
5
6 V
3.3 V, 60 Hz
3.273 to 3.327
6
6 V
3.3 V, 20 kHz
3.214 to 3.386
7
60 V
6.75 v, 60 Hz
6.68 to 6.82
8
60 V
33 V, 60 Hz
32.75 to 33.25
9
60 V
33 V, 20 kHz
N/A
10
60 V
33 V, 5 kHz
N/A
11
600 V
67.5 V, 60 Hz
66.8 to 68.2
12
600 V
330 V, 60 Hz
327.5 to 332.5
13
600 V
330 V, 2.5 kHz
323.0 to 337.0
14
1000 V
100 V, 60 Hz
97 to 103
15
1000 V
1000 V, 1 kHz
986 to 1014
16
AC Volts
Frequency
600 mV
150 mV, 99.95
kHz
99.93 to 99.97
17
600 mV
150 mV,
199.50 kHz
199.48 to 199.52
18
Sensitivity
6 V
0.7 V, 99.95
kHz
99.93 to 99.97
19
60 V
7 V, 99.95 kHz
99.93 to 99.97
20
Trigger level
6 V
3.4 V, 1 kHz
Sq. Wave
999.8 to 1000.2
21
Duty Cycle
6 V
5 V, 1 kHz, DC
offset 2.5 V
Sq. Wave
49.7 % to 50.3 %
22
DC Volts
6 V
3.3 V dc
3.297 to 3.303
23
60 V
33 V dc
32.97 to 33.03
24
600 V
330 V dc
329.7 to 330.3
25
1000 V
1000 V dc
998 to 1002