Fluke Corporation 8508A Extended Specifications
15
Notes:
[1] Specifications apply for max resolution in each function, normal mode
[2] Assumes 4 hour warm-up period
[3] Input zero or offset null required whenever the temperature moves more than ±1 °C from the temperature at which the
previous null/zero was performed
[4] TCal = Ambient calibration temperature
[5] Integration time > 1 Power Line cycle
[6] Valid for signals >1 % Full Scale, Transfer Mode On. Signal must be DC coupled <40 Hz. Readings invalid with Transfer Mode
On and 1 Hz filter selected when using internal trigger mode
[7] Max Volt.Hertz 3 x 10
7
[8] >300 V, <10 kHz add: ±0.0004 (R-300)
2
ppm
>300 V, 10 kHz - 30 kHz add: ± (0.0004 + (F - 10000)*1E-7)*(R-300)
2
ppm
>300 V, >30 kHz add: ±0.0024 (R-300)
2
ppm
[9] Typical below 10 Hz for ACV, below 10 Hz, and above 10 kHz for ACI and above 2 G
Ω
for Resistance
[10] Tru Ohms mode available on 2
Ω
to 20 k
Ω
ranges. Read Rate reduced in Tru Ohms mode. Specification for Tru Ohms same as
corresponding Normal or Lo Current range.
[11] Valid for 4-wire sensor
[12] Not including sensor uncertainty
[13] Assume Range and Full Scale = 2000 V when calculating for 1000 V Range. For DCI, additional errors only apply in 5 digit
resolution
[14] Fast mode not available in PRT
General specifications
General Specifications
Power
Voltage
100 V to 120 V ±10 %
or 200 V to 240 V ±10 %
Frequency
47 Hz to 63 Hz
Consumption
80 VA
Dimensions
Height
88 mm (3.5 inches)
Width
427 mm (16.8 inches)
Depth
487 mm (19.2 inches)
Weight
11.5 kg (25.5 lbs)
Environment
Temperature:
Operating
0 °C to 50 °C
Specified Operation
5 °C to 40 °C
Calibration (TCal)
20 °C to 25 °C
Factory Cal Temperature 23 °C
Storage
-20 °C to 70 °C
Warm Up
4 hours to full uncertainty
specification
Relative Humidity:
(non condensing)
Operating
< 90 % (5 °C to 40 °C)
Storage
< 95 % (0 °C to 70 °C)
Altitude:
Operating
< 2000 m
Storage
< 12000 m
Vibration and Shock:
Complies with MIL-PRF-28800F Class 3
EMC
EN50081-1 Class B, EN55011/22,
EN61326-1:1998, EN50082-1,
EN55011 1991 Class B,
EN61000-6-1:2001, FCC Rules part 15
sub part J class B, CE marked.
General Specifications (cont.)
Safety
Designed and tested to EN61010-1-2001,
UL 61010-1A1,
CAN/CSA 22.2 No. 61010.1,
CE and ETL marked.
Pollution Degree 2.
Installation Category II.
Equipment Class I (single insulation /
Earthed metal case).
Protection against water ingress IP4X
(general indoor conditions).
Input circuitry and connections
(creepage and clearances) designed to
interface at Installation or Measurment
Category I.
Measurement Isolation
Guard to Safety Ground
<3300 pF, >10 G
Ω
Lo to Guard:
In Remote Guard
<2800 pF, >10 G
Ω
(Not in Resistance function)
In Local Guard
Lo and Guard terminals are internally shorted
(in Resistance <2800 pF, >10 G
Ω
)
Autorange
Range Up
100 % of nominal range
Range Down
9 % of nominal range (18 % on 1 kV range)
Remote Interface
IEEE 488.2
Warranty
1 Year