Use this formula to determine the value of an external shunt resistor 
(meter impedance >10 M

Ω

):

Where:

Rs = Shunt resistance in M

Ω

Rm = Voltmeter input impedance in 

ΜΩ

 (>10 M

Ω

)

Example: If Rm = 20 M

Ω

,

Use this formula to calculate a correction factor (meter impedance 
<10 M

Ω

):

Where:

Cf = Correction factor (multiplier for meter reading) 

Rm = Voltmeter input impedance in M

Ω

Example: If Rm = 1 M

Ω

,

A meter reading of 0.526 volts represents an input of: 

0.526 x 1.901 = 1 or 1 kV

Circuit Loading 

The probe represents a load to the circuit:

1000 M

Ω

 , or 1

μ

A/1 kV, for 80K-40

75 M

Ω, 

or 13 

μ

A/1 kV, for 80K-6

Maintenance

The probe has no user-serviceable parts.
Use a soft cloth dampened with distilled water to clean the probe. Never 
use solvents of abrasive cleaners. Make sure the probes are dry before 
use.

Performance Test

Verify the accuracy of the probe with a voltage source measurement:

25 kV dc ±0.25 % for 80K-40

5 kV dc ±0.25 % for 80K-6

When used with a compatible dc voltmeter, the probe should measure the 
source with ±1 % accuracy. No calibration adjustments are available on 
the probe. 

Specifications

Voltage Range

80K-40:

0 kV to 

±

40 kV dc or peak ac, 28 kV rms ac

80K-6:

0 kV to 

±

6 kV dc or peak ac, 4.24 kV rms ac

Input Resistance

80K-40:

1000 M

Ω

80K-6:

75 M

Ω 

nominal

Division Ratio:

1000:1 (1000 x attenuation)

Rs

Rm x 10

Rm 10

=

Rs

20 10

×

20 10

-------------------

200

10

----------

20

 

M

Ω

=

=

=

Cf

1.11 +  Rm

1.11 x Rm

=

Cf

1.11 +  1

1.11 x 1

=

2.11

1.11

= 1.901

=