Theory Of Operation

The probe is a passive attenuator. See Table 1. High input impedance, 
accuracy, and stability characteristics are achieved with thick-film 
resistors. When the probe is connected to a voltmeter with a 10 M

Ω

 input 

resistance the probe becomes an accurate 1000:1 divider. This divider 
depends on a ground lead to complete the low side of the circuit path. The 
connection must always be secure before you do the voltage 
measurement. Otherwise, instrument damage or a shock hazard can 
result.

Table 1. Schematic Diagram

Voltmeter Compatibility

Accuracy of the meter is not included in the accuracy of the probe and 
must be added to the probe accuracy to determine system accuracy.
The probe is mechanically compatible with any ac or dc voltmeter or 
multimeter that accepts a standard double banana connector (spaced 
0.75" [19 mm]) with standard plugs (0.160" [4 mm]).
The probe is electrically compatible with any ac or dc voltmeter or 
multimeter that has an input impedance of: 

10 M

Ω

 

±

1 % for 80K-40

10 M

Ω

 

±

10 % for 80K-6

Voltmeters or multimeters with other input impedances require the use of 
an external shunt or a correction factor to obtain an accurate 
measurement. Higher impedance voltmeters or multimeters should be 
equipped with a shunt, and lower impedance voltmeters or multimeters 
should be assigned correction factors. 

Probe

Resistor

RA

RB

80K-40

999 M

Ω

 1.11 

M

Ω

 

80K-6

75 M

Ω

 75 

k

Ω

 

Probe Tip

(Dual Banana Plug)

Protective

Earth Lead

RA

RB

10 MΩ
(Voltmeter)

HI

LO